User's Manual Part 10

Directions for Use Specifications 175
Patient leakage current 10 μA maximum, normal condition type CF per IEC EN 60601-1
50 μA maximum, single-fault condition type CF per IEC EN 60601-1
Enclosure leakage current 100 μA maximum, per IEC EN 60601-1
Protection against harmful
ingress of water
IPX1 when the LED faces up
IPX0 otherwise
Humidity 95% per MILT-28800 Par 4.5.5.1
Strain-Relief Strength 30 N per UL/IEC EN 60950-1 and UL/IEC EN 60601
Weight 0.8 lb
Insulation class (transformer) UL Rated 105° C
Width 2.68 in (68 mm)
Height 1.54 in (39 mm)
Length 4.13 in (105 mm)
Weight 10.6 oz (300 g)
AC input connector IEC 60320
-C8, 2.5 A rated input housing
DC output connector Hypertronics DO2
-size latching 3-pin male
Output cable length 8 feet for IEC power cord
Table 25. Specification: Monitor Battery
Characteristic Specification
Type 7.4 V, rechargeable, lithium ion. Field
-replaceable by service technician.
Capacity > 1800 mA
-hr minimum
Charger Contained in the cradle. Batteries are charged in place in the monitor.
Fuse rating 5A, 125V – User cannot access the fuse.
Note: Internal electronic overload circuitry is used as the primary method of protection. This
circuit resets itself when an overload is removed.
Recharge time 3 hours at 25° C (typical), with monitor either on or off.
Storage < 30 days
-4°F to 122°F -20°C to 50°C
30 days to 90 days
-4°F to 104°F -20°C to 40°C
> 90 days
-4°F to 86°F -20°C to 30°C
NOTE: For best charge retention during extended storage, remove the battery from the monitor.
Cycle life 300 full charge/discharge cycles @ 20° C (to 70% of initial capacity)
Table 24. Specification: Power Adapter (continued)