Guidance and manufacturer’s declaration—electromagnetic emissions
The battery-operated Propaq LT Series monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the monitor should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment—guidance
RF emissions
CISPR 11
Group 1 The battery-operated Propaq LT Series monitor uses RF
energy only for its internal function.
a
Therefore, its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
a. The battery-operated Propaq LT Series monitor contains a 2.4-GHz frequency-hopping spread-spectrum transmitter for the purpose
of wireless communication. The radio is operated according to the requirements of various agencies, including FCC 47 CFR 15.247 and
R&TTE Directive (1995/5/EC). The radio is excluded from the EMC requirements of 60601
-1-2:2001, but should be considered when
addressing possible interference issues between this and other devices.
RF emissions
CISPR 11
Class B The battery-operated Propaq LT Series monitor is suitable for
use in all establishments, including domestic establishments
and those directly connected to the public low
-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
No connection to mains
(battery-operated)
Voltage fluctuations/flicker
emissions
IEC 61000
-3-3
No connection to mains
(battery