User's Manual Part 10

A
163
Specifications
ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .163
Resp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .166
SpO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .167
NIBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .168
Alarms and Alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .170
Wireless Monitor (Model 802LTRN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .173
Physical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .173
Power. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174
Cradle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176
ECG
Table 13. Specification: ECG
Characteristic Specification
Regulatory standards Meets ANSI/AAMI EC13-2002 and EN 60601-1:1990 + A1 + A2
Connector Hypertronics 9-pin female D01 latching connector
Selectable leads
3-lead cable
5-lead cable
Displayable I, II or III
Internally provides I, II or III
Displayable I, II, III, aV
R
, aV
L
, aV
F,
V
Internally provides I/II, III and V
Cable type detection Automatic detection and differentiation of 3
- or 5-lead cable
Lead fault indicator Displayed chest icon with flashing indicator for each electrode
ECG Size (sensitivity) 0.2, 0.5, 1, 2, 4, and 8 mV/cm
Display sweep speed 25 mm/sec
Bandwidth (monitor display) 0.5 to 40 Hz Monitor Mode, Adult
0.05 to 40 Hz Extended Mode, Adult
0.5 to 80 Hz Monitor Mode, Pediatric or Neonate
0.05 to 80 Hz Extended Mode, Pediatric or Neonate
Bandwidth (to Acuity) 0.05 to 80 Hz Monitor or Extended Mode, Adult, Pediatric, or Neonate
Notch filters 50 and 100 Hz, 60 and 120 Hz, or OFF, selectable at Acuity or at the monitor