Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02 Autozero failure: Check for air obstruction; limit patient movement.
C03 Inflation too rapid: Check for kinked hose or air obstruction.
C04 Excessive inflation time: Check for air leaks.
C05 Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06 Measurement outside of device’s range: Check patient condition.
E10 Cuff overpressure: Check patient condition.
Temperature Error Codes
C20 Broken/missing probe: Replace probe.
P Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6 SpO
2
heart rate out of range.
C8 Bad sensor: Replace sensor.
E7 Internal malfunction: Contact Technical Service.
C9 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11 Internal safety violation: Check patient; contact Technical Service.
C12 Ambient temperature out of range: Adjust room temperature or device location.
C13 Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.
For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.
Reorder 4200-88E
706287 Ver. B
Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02 Autozero failure: Check for air obstruction; limit patient movement.
C03 Inflation too rapid: Check for kinked hose or air obstruction.
C04 Excessive inflation time: Check for air leaks.
C05 Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06 Measurement outside of device’s range: Check patient condition.
E10 Cuff overpressure: Check patient condition.
Temperature Error Codes
C20 Broken/missing probe: Replace probe.
P Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6 SpO
2
heart rate out of range.
C8 Bad sensor: Replace sensor.
E7 Internal malfunction: Contact Technical Service.
C9 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11 Internal safety violation: Check patient; contact Technical Service.
C12 Ambient temperature out of range: Adjust room temperature or device location.
C13 Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.
For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.
Reorder 4200-88E
706287 Ver. B
Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02 Autozero failure: Check for air obstruction; limit patient movement.
C03 Inflation too rapid: Check for kinked hose or air obstruction.
C04 Excessive inflation time: Check for air leaks.
C05 Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06 Measurement outside of device’s range: Check patient condition.
E10 Cuff overpressure: Check patient condition.
Temperature Error Codes
C20 Broken/missing probe: Replace probe.
P Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6 SpO
2
heart rate out of range.
C8 Bad sensor: Replace sensor.
E7 Internal malfunction: Contact Technical Service.
C9 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11 Internal safety violation: Check patient; contact Technical Service.
C12 Ambient temperature out of range: Adjust room temperature or device location.
C13 Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.
For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.
Reorder 4200-88E
706287 Ver. B
Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02 Autozero failure: Check for air obstruction; limit patient movement.
C03 Inflation too rapid: Check for kinked hose or air obstruction.
C04 Excessive inflation time: Check for air leaks.
C05 Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06 Measurement outside of device’s range: Check patient condition.
E10 Cuff overpressure: Check patient condition.
Temperature Error Codes
C20 Broken/missing probe: Replace probe.
P Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6 SpO
2
heart rate out of range.
C8 Bad sensor: Replace sensor.
E7 Internal malfunction: Contact Technical Service.
C9 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11 Internal safety violation: Check patient; contact Technical Service.
C12 Ambient temperature out of range: Adjust room temperature or device location.
C13 Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.
For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.
Reorder 4200-88E
706287 Ver. B
Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02 Autozero failure: Check for air obstruction; limit patient movement.
C03 Inflation too rapid: Check for kinked hose or air obstruction.
C04 Excessive inflation time: Check for air leaks.
C05 Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06 Measurement outside of device’s range: Check patient condition.
E10 Cuff overpressure: Check patient condition.
Temperature Error Codes
C20 Broken/missing probe: Replace probe.
P Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6 SpO
2
heart rate out of range.
C8 Bad sensor: Replace sensor.
E7 Internal malfunction: Contact Technical Service.
C9 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11 Internal safety violation: Check patient; contact Technical Service.
C12 Ambient temperature out of range: Adjust room temperature or device location.
C13 Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.
For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.
Reorder 4200-88E
706287 Ver. B
Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02 Autozero failure: Check for air obstruction; limit patient movement.
C03 Inflation too rapid: Check for kinked hose or air obstruction.
C04 Excessive inflation time: Check for air leaks.
C05 Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06 Measurement outside of device’s range: Check patient condition.
E10 Cuff overpressure: Check patient condition.
Temperature Error Codes
C20 Broken/missing probe: Replace probe.
P Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6 SpO
2
heart rate out of range.
C8 Bad sensor: Replace sensor.
E7 Internal malfunction: Contact Technical Service.
C9 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11 Internal safety violation: Check patient; contact Technical Service.
C12 Ambient temperature out of range: Adjust room temperature or device location.
C13 Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.
For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.
Reorder 4200-88E
706287Ver. B
Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02 Autozero failure: Check for air obstruction; limit patient movement.
C03 Inflation too rapid: Check for kinked hose or air obstruction.
C04 Excessive inflation time: Check for air leaks.
C05 Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06 Measurement outside of device’s range: Check patient condition.
E10 Cuff overpressure: Check patient condition.
Temperature Error Codes
C20 Broken/missing probe: Replace probe.
P Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6 SpO
2
heart rate out of range.
C8 Bad sensor: Replace sensor.
E7 Internal malfunction: Contact Technical Service.
C9 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11 Internal safety violation: Check patient; contact Technical Service.
C12 Ambient temperature out of range: Adjust room temperature or device location.
C13 Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.
For technical questions inside the U.S., please call Welch Allyn at 800-535-6663.
Outside the U.S., please contact your local distributor.
Reorder 4200-88E
706287Ver. B
Error Codes
When responding to an error code, always check the patient first and then the device.
Blood Pressure Error Codes
C02 Autozero failure: Check for air obstruction; limit patient movement.
C03 Inflation too rapid: Check for kinked hose or air obstruction.
C04 Excessive inflation time: Check for air leaks.
C05 Excessive noise: Check patient condition, cuff placement; limit patient movement.
C06 Measurement outside of device’s range: Check patient condition.
E10 Cuff overpressure: Check patient condition.
Temperature Error Codes
C20 Broken/missing probe: Replace probe.
P Loss of tissue contact: Ensure proper probe positioning.
E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location.
C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
SpO
2
Error Codes
C6 SpO
2
heart rate out of range.
C8 Bad sensor: Replace sensor.
E7 Internal malfunction: Contact Technical Service.
C9 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.
Verify Spot Vital Signs is not being used for monitoring purposes.
General Error Codes
E11 Internal safety violation: Check patient; contact Technical Service.
C12 Ambient temperature out of range: Adjust room temperature or device location.
C13 Battery Depletion: Use wall transformer.
E0.0-E9.9 Temperature module malfunction: Contact Technical Service.
E20-E50 General internal malfunction: Contact Technical Service.