18 Chapter 1 General information Welch Allyn Micropaq Monitor
5. Press to highlight SERVICE MENU, then press to display the Service Menu.
6. Press to highlight DEMO MENU, then press to display the Demo Menu.
7. Press to highlight DEMO 1 or DEMO 2, then press to start.
8. While in Demo mode you can practice changing settings such as ECG lead selection
and alarm limit adjustment. (These changes only affect the Demo mode and are
erased when you exit the Demo mode.)
9. To change to the other Demo selection, press to display the menu, then scroll
down to highlight TOGGLE DEMO MODE and press .
10. To exit the Demo mode, either insert a patient cable or remove and insert the battery.
The monitor restarts and enters the normal monitoring mode.
Demo Mode Display Values and Alarm Limits
Display Demo 1 Demo 2
a
a. Demo 2 will cause patient alarms.
Alarm Limits (On)
ECG Waveform Normal sinus rhythm,
normal ST
Normal sinus rhythm,
normal ST
(not applicable)
ECG Heart Rate
SpO
2
Pulse Rate
80 125 Lower 50
Upper 120
SpO
2
Saturation% 97 88 Lower 90
Upper 100
Service Menu
A simulated waveform and
numerics are displayed.
Demonstration Mode
SIMULATION indicates the
Demo mode is active.