User's Manual Part 2
Directions for Use Chapter 5 Maintenance 45
Inspect and clean the monitor and accessories
Before cleaning, thoroughly inspect the monitor and all accessories for any signs of
damage, cracks, or improper mechanical function of keys or connectors. While gently
bending and flexing cables, inspect for damage, cracks, cuts, abrasions, extreme wear,
exposed wires or bent connectors. Confirm connectors securely engage. Report damage
or improper function to your service department. At least every 12 months, be sure to
thoroughly inspect the battery charger case and power cord for damage or extreme wear.
To clean the monitor, batteries, or battery charger:
1. Wipe the equipment with a nearly dry cloth moistened with one of the approved
cleaning solutions listed in the table on “Approved Cleaning Solutions” on page 46.
2. After cleaning, thoroughly wipe off any excess cleaning solution with a soft cloth
dampened with water, then dry. Do not let the cleaning solution run into or
accumulate in connector openings, latches, or crevices. If liquid gets into a battery
well or connectors, dry the area with warm air, then check the equipment to confirm
that it operates properly.
WARNING Be sure to unplug the monitor battery charger power cord from the
electrical power outlet before inspecting or cleaning the battery charger.
Exposing the battery charger to liquids such as cleaning solutions while
connected to electrical power could result in electrical shock or fire.
WARNING Do not autoclave the monitor, battery, battery charger, or
accessories. Do not immerse the monitor, battery, or battery charger in liquid
when cleaning. Do not immerse accessories in liquid when cleaning unless the
accessory manufacturer’s cleaning instructions explicitly instruct you to do so.
Caution Use only cleaning solutions which are recommended by Welch Allyn
for this equipment. Use of solutions which are not recommended or which have
a high acid content or are otherwise inappropriate can cause damage to the
equipment, including cracking and deterioration of the plastic case. Do not use
these solutions or similar products: Butyl alcohol, Denatured ethanol, Freon
™
,
Mild chlorine bleach solution, Isopropyl alcohol (except for the SpO
2
cable),
Trichloroethane, Trichloroethylene, Acetone, Vesphene II, Enviroquat
®
,
Staphene
®
, Misty
®
, Virex
®
, Glutaraldehyde, Formula 409
®
, or Fantastik
®
.