User's Manual Part 2

6
49
Reference
Operating settings
The following table lists all of the monitor settings and the default settings.
Parameter Set at
Monitor
Set at Acuity Previous
Setting
Retained at
Monitor at
Power-Up
Monitor Default
Setting
For Each
Patient
a
a. Set by clinician at Acuity Central Station.
For Entire
Acuity
Unit
b
b. Set by Acuity System Administrator during system installation.
Patient Mode
Adult (age 13 years and older)
Pediatric (age greater than 28
days of age or more than 44
weeks gestation up to 12
years)
No Yes No Yes Adult
ECG screen mode (Single, Dual, 5
Sec, or Full Screen)
Yes No Yes Yes Single
ECG 1 Lead Selection Yes Yes Yes Yes II
ECG 2 Lead Selection Yes Yes Yes Yes V (or III if no V lead)
ECG Size (Scale) Yes No No Yes 1 mV/cm
Language No No Yes Yes English
Mains Filter (off, 50, or 60 Hz) No No Yes Yes 60
Vital Signs Alarm Volume
(high, low, or off)
With Acuity Connection
Without Acuity Connection
No
No
No
No
Yes
Yes
No
Yes
off
high
Equipment Alert Volume
(high, low, or off)
With Acuity Connection
Without Acuity Connection
No
No
No
No
Yes
Yes
No
Yes
off
low
HR/PR Alarm Limits
(Lower, Upper)
Yes Yes Yes Yes Adult: 50, 120 bpm
Ped: 50, 150 bpm
SpO
2
Alarm Limits
c
(Lower, Upper)
c. SpO
2
alarm limit range depends on the software version of the Acuity System to which the monitor is connected. (See “Heart rate and
arrhythmia analysis option” on page 53 and “Pulse oximetry (SpO
2
) specifications - Nellcor” on page 55.)
Yes Yes Yes Yes Adult: 90, 100%
Ped: 90, 100%
Regulatory settings
(U.S., Europe, Japan)
No No Yes Yes U.S.
Pacer Detection Enable No Yes No Yes On
Menu Lockout No No Yes No Off
Display Backlight Timeout No No Yes Yes 120 seconds