User's Manual Part 2

58 Chapter 6 Reference Welch Allyn Micropaq Monitor
Physical specifications
Battery specifications
Protection classifications, all configurations
Characteristic Specification
Type of Protection against Electric
Shock—Monitor Type: CF
Battery operation only
Battery must be recharged in separate battery charger.
IEC EN 60601-1, 2nd Edition
Degree of Protection Against Electric
Shock, for Parts Applied to Patients
See monitor labels. CF defibrillator protected.
IEC EN 60601-1, 2nd Edition
Recovery time following defibrillator
discharge
Less than or equal to 10 seconds
Method of Disinfection Not suitable for autoclaving (see cleaning instructions on “Inspect and clean
the monitor and accessories” on page 45).
Flammable Anesthetics Not suitable for use with flammable anesthetics.
Height 7.80 in (19.8 cm)
Width 3.50 in (8.9 cm)
Depth 1.96 in (4.9 cm)
Weight (including battery)
Model 406
Model 408
17.0 oz (0.48 kg)
18.6 oz (0.53 kg)
Characteristic Specifications
Reorder Number
Lithium Ion Battery
2EA Active A
008-0647-XX
Battery Type Rechargeable, Lithium Ion
Battery Capacity 2 cells, 7.4 V (nominal), 8.4 V (charging), 1800 mA-hr
Battery Weight 4.5 oz (0.13 kg)
Battery Charger External device
Battery Fuse Rating 5 A, 125 V (not user-accessible)
Note: Internal electronic overload circuitry is used as
the primary method of protection.
Operating Times on Battery
a
a. Battery operating times based on these conditions: new fully-charged battery operating at 25° C,
the monitor connected to Acuity, eight patient alarms per hour, minimal motion artifact.
Monitor w/ ECG only: 25 hrs
Monitor w/ ECG and SpO
2
: 10 hrs
Battery Recharge Time 4 hours at 25° C (typical)
Battery Lifetime 300 charge/discharge cycles to 70% of original
capacity (typical)
Note
The following factors may reduce battery operating time:
Amount of time not connected to Acuity.
Frequency and duration of alarms and alerts.
Amount of operator activity using monitor keys (activates display).
Age of battery.
Amount of motion artifact during SpO
2
monitoring.
WARNING The monitor battery is Lithium Ion. Do not incinerate, submerge,
crush, disassemble, or autoclave. If a battery has been submerged in liquid,
discard the battery properly; do not try to recharge or reuse the battery. Do not
short the battery terminals. Do not try to connect the battery to any device
except the monitor or the battery charger. Do not expose to high temperature
(above 60° C or 140° F). Use only the specified monitor battery charger.