DataVU 5 - Operating Manual 59480
Contents 1 Introduction 7 1.1 Preface .......................................................................................................... 7 1.2 Arrangement of the documentation ........................................................... 8 1.2.1 Structure of this operating manual ................................................................. 8 1.3 1.3.1 1.3.2 1.3.3 Typographical conventions ......................................................................... 9 Warning signs .............
Contents 3.2 3.2.1 3.2.2 3.2.3 3.2.4 3.2.5 3.2.6 3.2.7 Visualization ................................................................................................ Diagram representation with digital display (small measurements) ............. Diagram representation with scaling ............................................................ Diagram representation with bar graph ........................................................ Large digital display (large measurements) ...............................
Contents 5.3 User interface ............................................................................................. 79 5.4 Configuration .............................................................................................. 81 5.5 Teleservice .................................................................................................. 84 5.6 Data transfer from and to the instrument ................................................ 86 5.6.1 Transfer via CompactFlash memory card ......
Contents 10 Electrical connection 121 10.1 Installation notes ...................................................................................... 121 10.2 Technical data ........................................................................................... 121 10.3 Connection diagram ................................................................................
1 Introduction 1.1 Preface B Please read these operating instructions before commissioning the instrument. Keep the operating instructions in a place that is accessible to all users at all times. Please assist us to improve these operating instructions, where necessary. Your comments will be appreciated. H All necessary settings are described in this manual.
1 Introduction 1.2 Arrangement of the documentation The documentation for this instrument consists of the following parts: Operating Manual 59480 This operating manual comes with the delivery. It is addressed to the equipment manufacturer and the user with appropriate technical expertise. Interface Description 59482 It provides information on the serial interface (RS232 and RS485), which can be supplied as an extra code.
1 Introduction 1.3 Typographical conventions 1.3.
1 Introduction 1.3.3 Representation modes Keys h+E Keys are shown in a box. Both symbols and text are possible. If a key has a multiple function, then the text shown is the one that corresponds to the function that is active at the moment. Screen texts Program Manager Texts that are displayed in the setup program are indicated by italic script. Menu items Edit Device data 10 Menu items in the setup program which are referred to in this manual are shown in italics.
2 Device description 2.1 Displays and controls Color display 320 x 240 pixels, 27 colors Menu-dependent function keys (softkeys) screen-dependent functions, represented by text or symbols Status LED (red) is on continuously while an alarm is present Power LED (green) This is lit continuously as soon as power is applied. It blinks when the screen saver is active.
2 Device description Opening and closing the door of the instrument Header Status line To open or close the door of the instrument, turn the knob 2.2 Operating principle and graphic elements Keys The recorder is operated from eight keys. Three of these keys have fixed functions, the other five (softkeys) have menu-dependent functions. v Chapter 2.1 “Displays and controls” Softkeys 12 The functions of the softkeys appear in the bottom line of the display, as symbols or in plain text.
2 Device description Status line The status line is shown in the top section of the display. It provides information on important actions and states. The status line is always visible, irrespective of the level (operation, parameters, configuration). Alarm If an alarm occurs (e.g. out-of-limit), the bell (alarm) will flash in this field. CF card / internal memory Shows the free storage of the CompactFlash card or the internal memory, as a percentage.
2 Device description Channel line (channel representation) The channel line shows the measurements of the active channels and their unit as - a measurement - scale or - bar graph. Alternatively, the header can be switched off altogether. In addition, alarms and out-of-limit conditions are made directly visible, depending on the display mode. Example: Measurement (small meas.) Numerical display The measurements are shown in numerical form.
2 Device description 2.
2 Device description Representation Representation on the screen is as follows: Representation as symbol On/off represented as switch: as diagram Representation as a record over time: Outputs The digital signals can be used to operate three relays (extra code). The action can be configured as a break (SPST-NC) or make (SPST-NO) contact (Configuration ➔ Outputs). External texts External texts can be called up via four binary inputs or the logic channels.
2 Device description Counters Counter inputs Any of the digital signals can be used as input signals for the counters. v Chapter 2.4 “Digital signals (event traces)” Example: - binary inputs - logic channels - alarms - memory alarms - Modbus flag (signal via interface) Counter frequency 30Hz max. Evaluation The count pulses can be evaluated (weighted). A down counter can be implemented by entering a negative weighting (e.g. weighting factor -1).
2 Device description Operating time counter The operating time counter will count how long a selected binary input or one of the digital signals is closed (set). The time can be displayed in sec, min, hr or days. 2.5.1 Reporting period of the counts After an adjustable time (reporting period), the counts are stored for all counters/integrators/operating time counters. The counts for the current and the previous 7 completed reporting periods will be displayed.
2 Device description External reset You can configure one control signal for all 6 counters/integrators together, with the result that the counters/integrators are reset to 0 without any saving of the previous values. The period for the counter/integrator summation will be restarted at this point. This means that after the test run of an installation, for instance, the recording can be freshly started; the test run values that are not required are eliminated.
2 Device description 2.5.4 Response at end of summer time A If a period of less than one hour has been set for counters/ integrators, and the change from summer time to winter time falls precisely within this period, then the PC evaluation software (PCA3000) will forcibly set the end time to be the same as the start time. This prevents a negative period arising, and ensures that further processing is error-free. 2.6 Math/logic module The math and logic module is available as an extra code.
2 Device description The following variables are used for the formulae: - analog inputs (AE1 — AE6) - math channels (MAT1 — MAT6) - counter/integrator channels (ZI1 — ZI6) - binary inputs (BE1 — BE4) - alarms - errors - Modbus flag (signal via interface) - instrument-specific data (only after consultation with the manufacturer) If counter/integrator values are used for calculation, then please note that their accuracy is reduced, since, in this case, two different data formats have to be used for calculatio
2 Device description Logic module You also have up to 6 channels available as logic channels. As is the case with all the other digital signals, the calculated digital (Boolean) values can be used for different functions: - recording in the event traces, as a control signal for display switch-off, clock time synchronization, operating time counter, externally controlled counters / integrators counter/integr. reset event operation and key inhibit, for output to a relay and as count input for a counter.
2 Device description 2.7 Operating modes 3 operating modes The instrument has 3 operating modes: - normal operation - timed operation - event operation The following settings can, among others, be made for each of the three operating modes: - stored value - storage cycle Stored value The stored value setting determines what is stored: average, minimum, maximum or momentary value between two storage cycles, or the peak value (envelope).
2 Device description 2.8 Storing data Life-cycle data management The integrated life-cycle data management puts users in the position of being able to save all the process data from the system that is being monitored in an archive file on their PCs or a server system.
2 Device description Operating principle The measurements from the analog inputs are acquired continuously in a 250msec sampling cycle, and then stored in the main memory. These measurements also serve as the basis for limit monitoring. Main memory (FLASH memory) The data stored in the main memory are regularly copied to the CompactFlash card in 4 kbyte blocks. The main memory is written to as a ring memory, i.e. when it is full, the oldest data will automatically be overwritten by new data.
2 Device description Recording duration The recording duration depends on various factors: - number of analog channels and event traces being recorded - storage cycle - number of events in the event list. Internal recording duration (without external CF card) Number of channels Memory size Storage cycle 1 min Storage cycle 30 sec Storage cycle 10 sec Storage cycle 1 sec 3 approx. 1MB 42.2 days 21.1 days 7 days 17 hours 6 approx. 1MB 29.5 days 14.8 days 4.
2 Device description 2.9 Read-out of data In addition to automatic read-out via the external CompactFlash memory card, measurement data can also be read out through one of the interfaces (RS232, RS485, setup, Ethernet). Both read-out options (card/interface) work in parallel. For this reason, there are also two “digital signals”, which indicate when the available storage space has fallen below a certain configurable value.
2 Device description 2.11 Webserver Event traces Counters / integrators Analog channels Device name If the paperless recorder is equipped with the extra code “Ethernet interface”, and the device-internal Ethernet software (“Ethernet version“) has the number 183.02.03 or higher, then the paperless recorder has an inbuilt webserver. The webserver can be started by entering the IP address in an Internet browser on the PC (e.g. http://10.10.90.45).
2 Device description Display of ... Activated through menu ... Device name Configuration - Instrument data - Device name Analog or math channels Configuration - Recording - Analog channels Counters or integrators Configuration - Counter/integrator Event traces Configuration - Recording - Event traces Only the activated signals will be displayed. The information about the “Ethernet version” can be found in the “Device info” menu. v Chapter 3.
2 Device description 30
3 Operation and visualization After starting up the paperless recorder by switching on the supply (power ON), you will see the start logo. During screen build-up, the recorder is initialized with the data of the last configuration. H H After the initialization phase, the measurement diagram (visualization level) is displayed. By using the setup program, an individual start logo (bitmap with a maximum of 16 colors) can be loaded into the instrument, through the function Extras Start screen.
3 Operation and visualization 3.1 Basic menu The basic menu is the central point from which the various instrument levels branch out. The following levels are available: v Chapter 3.2 “Visualization” v Chapter 3.3 “Parameterization” v Chapter 3.4 “Configuration” v Chapter 3.5 “Event list” v Chapter 3.6 “CompactFlash card” v Chapter 3.7 “Device info” h Select the required level h Enter selection with E The basic menu is displayed after pressing the Mkey1.
3 Operation and visualization 3.2 Visualization As already mentioned in Chapter 2.2 “Operating principle and graphic elements”, the softkeys are arranged at the bottom of the screen. They change their function according to the menu and are shown as symbols or plain text.
3 Operation and visualization 3.2.1 Diagram representation with digital display (small measurements) Diagram representation can be reached from the basic menu, by calling up the “Visualization” menu or by pressing the D key.
3 Operation and visualization 3.2.2 Diagram representation with scaling Scaling start of the selected channel Lower limit index of the selected channel (no display when alarm is off) Current measurement Channel name Upper limit index (no display when alarm is off) Scaling end of the selected channel Which scaling (on which channel) is to be displayed can be selected through the parameter Parameterization ➔ Diagram view ➔ Channel indication. 3.2.
3 Operation and visualization 3.2.4 Large digital display (large measurements) This display type is limited to the digital display. If more than 6 signals are active (for instance, 6 analog inputs and 3 math channels), the button is switched into the display. It can be used to switch between the individual signals.
3 Operation and visualization 3.2.5 Curve representation (header switched off) This display type is limited to the curve display.
3 Operation and visualization 3.2.6 Evaluation of the stored measurement data History The softkey function changes during evaluation. In addition, the current zoom factor and cursor position (date and time) will be shown.
3 Operation and visualization A Zoom Scrolling backwards within the internally stored measurement data is possible up to the last changeover from summer time to normal time. If the zoom factor has to be adjusted, or specific times have to be searched, then it is necessary to switch the softkey functions. h Press softkey The degree of compression of the measurement data on the screen is given as a ratio in steps (1:1, 1:2, 1:5, 1:10, 1:20, 1:50 and 1:100).
3 Operation and visualization Cursor positioning After activating the button, the following dialog appears for positioning the cursor on a specific time: After entering the date and time and activating positioned on the selected time. , the cursor is If no measurement data have been stored for the selected time, then the cursor is positioned on the next possible time.
3 Operation and visualization 3.2.7 Counters / integrators / operating time Count display is optionally available (extra code). If it is available, a screen such as the following will appear after the softkey has been pressed, in which the current counts are shown in a numerical form. Show curves Expanded representation Programmed alarm limit was exceeded Overrange Press the softkey to activate the expanded representation.
3 Operation and visualization Expanded representation Normal representation Current count Concluded reporting period Show next concluded reporting period The last 7 concluded reporting periods can be shown, one after another, by pressing the key. 42 Event list see Chapter 3.
3 Operation and visualization 3.3 Parameterization The Parameter level is available to set - contrast, - speed indication - display off - diagram view and - counter/integrator reset. H All parameters are selected using the or and and keys. Contrast The contrast of the screen can be set here. This ensures that the screen is always highly legible, even under difficult light conditions. Speed indication “mm/h”, “time/div” or “storage cycle” is selected here for the speed display in the diagram.
3 Operation and visualization Display off (screen saving) Switch-off event = waiting time For screen saving, a time between 0 and 32,767min can be set under the parameter “Waiting time”. If no key on the recorder is operated during this time, then the screen goes dark. The power LED blinks during screen saving. Screen saving can be canceled by pressing any key on the recorder. If 0min is set, then screen saving is deactivated.
3 Operation and visualization Counter/ integrator reset After the password has been successfully entered, the counts for each of the 6 channels can be set to 0 or a defined value in this menu. When a value has been entered (confirm with the E key), a message with the new and old count is entered in the event list. The time period for the counter/ integrator summation will not be restarted. The previous counter/integrator values will also not be saved.
3 Operation and visualization 3.4 Configuration When you call up the configuration level, you will be asked to enter the password (factory setting: 09200). This also serves to prevent unauthorized alteration of the configuration. After entering the password, acknowledged with the E key. a warning appears, which must be v Chapter 3.10 “Code number (password query)” Window technology As for the other levels, the configuration principle is also based on menu-led window technology.
3 Operation and visualization The configuration of the recorder is sub-divided into the following levels: v Chapter 4 “Configuration parameters” 3.5 Event list Events Various events can initiate texts which are incorporated in the event list and saved to the internal memory or CF card. Events may include: - alarms triggered by out-of-limit conditions on individual channels, - external texts triggered through binary inputs or logic channels, - system messages (e.g.
3 Operation and visualization Standard texts The recorder offers standard texts as listed in the following table: Standard text Comment Channelx low alarm ON Channelx low alarm OFF Channelx high alarm ON Channelx high alarm OFF Counter/int. alarm x ON Counter/int.
3 Operation and visualization Basic menu ➔ Event list The event list is called up via the basic menu: h Select operating level Event list h Enter selection with E Event list jump back directly into diagram presentation 49
3 Operation and visualization 3.6 CompactFlash card Automatic storage of measurement data The data stored in the measurement data memory (FLASH) of the paperless recorder are saved at regular intervals to the CompactFlash card1 in the instrument. The PC evaluation program (v Chapter 7 “PC programs”) reads the data from the CF card and provides user-friendly functions for evaluation.
3 Operation and visualization CompactFlash card Measurement data not yet saved are written to the CF card. Counter/integrator reporting periods are concluded and written to the CF card, together with the measurement data not yet saved. Counters/integrators are reset (to 0) and restarted.
3 Operation and visualization Status messages 52 Status messages from the CompactFlash manager are shown in the corresponding action window. The following status messages are possible: Status message Description CF CARD UPDATED Directly before removing the CF card from the instrument, it is necessary to call up Update CF card so that all measurement data up to the time of removal are contained on the CF card. The data not yet stored since the last automatic saving are written to the CF card.
3 Operation and visualization 3.7 Device info The device info window displays general information about the instrument. It also includes the errors “Battery empty” and “Data lost”. If one of the instrument errors occur, the info symbol flashes in the status line.
3 Operation and visualization - This menu also includes the version number (e.g. 208.01.04). This is the version number of the device software. It is important because some functions are only available from a certain version number onwards. - If the “Digital I/O option” is available (Yes), then the recorder also includes an RS232/RS485 interface. - If the Ethernet interface is available, then the field “Vers. No. Ethernet” includes a number combination (e.g. 183.02.02).
3 Operation and visualization 3.8 Text entry Entry options The configurable texts can be entered either through the setup program or on the instrument itself. This section describes the entry on the instrument. Character selection The display below is shown when a text (e. g. Configuration ➔ Texts) has been selected at the configuration level for editing using E.
3 Operation and visualization 3.9 Entering values The softkeys are also used to enter values in the instrument – the example shows fine calibration being carried out. Increase (+1) or decrease (-1) selected digit or change the sign H Select the digit to be altered The sign can only be altered if the value is not “0”. This means that you cannot start the value entry with an alteration of the sign if the current value is “0”. The entry can be concluded with E or canceled with D.
3 Operation and visualization 3.10 Code number (password query) The following functions are protected ex-factory against unauthorized access by a password query: - the Configuration menu - parts of the CompactFlash card menu - the Parameterization ➔ Reset counter / int. menu. The factory default setting for the passwords is 09200. You can also use a password to secure access to the recorder via one of the interfaces (with the exception of the setup interface).
3 Operation and visualization Password query in the configuration menu After the password has been entered in the Configuration menu, an additional security query will appear. You will only be able to gain access to the parameters when you have confirmed the the query with E.
4 Configuration parameters 4.
4 Configuration parameters 4.2 Table of configuration parameters The table below lists all the instrument parameters. The order in which the parameters are explained corresponds to the order in which they appear on the instrument (in the menu structure). The first column describes the path to the particular parameter, via the menus and windows. The second column lists the possible settings for the parameters or the possible selections. The factory default setting in this column is shown bold.
4 Configuration parameters Parameter Value/Selection Description Channel representation Parameterization ➔ Diagram view ➔ Chan. represent. Meas. small, Scaling Bar graph Meas.
4 Configuration parameters 4.2.2 Configuration - Instrument (device) data Parameter Value/Selection Device name Configuration ➔ Device data ➔ Device name 16 characters Date Configuration ➔ Device data ➔ Date and time ➔ Date any date Entry of the current date Time Configuration ➔ Device data ➔ Date and time ➔ Time any time Entry of the current time Time synchronization Configuration ➔ Device data ➔ Date and time ➔ Time synchroniz. Off, Bin. inp.
4 Configuration parameters Parameter Value/Selection Language Configuration ➔ Device data ➔ Language German, English, French, Dutch, Spanish, Italian, Hungarian, Czech, Swedish, Polish, Danish, Finnish, Portuguese, Russian Temperature unit Configuration ➔ Device data ➔ Temperature unit °C, °F Memory upload Configuration ➔ Device data ➔ Memory upload via CF card, via RSxxx You can determine how data are mainly to be uploaded.
4 Configuration parameters Parameter Enable options Configuration ➔ Device data ➔ Enable options Enable options: Computed code No. Configuration ➔ Device data ➔ Enable options ➔ Computed code No. Enable options: Configuration Authorization code ➔ Device data ➔ Enable options ➔ Authorization code Device-info text Value/Selection Description This parameter is used to enable additional functions, e.g. the math/ logic module and counter/ integrator.
4 Configuration parameters Parameter Value/Selection Description External cold junction temperature Configuration ➔ Analog inputs ➔ Analog input 1—6 ➔ Ext. CJ temp.
4 Configuration parameters 4.2.4 Configuration - Math channels Parameter Value/Selection Description Scaling start Configuration ➔ Math channels ➔ Math channel 1 — 6 ➔ Scaling start -99999 to 0 to +99999 Scaling start of a math channel. The setup progam is used to determine what is involved in a math channel. Scaling end Configuration ➔ Math channels ➔ Math channel 1 — 6 ➔ Scaling end -99999 to +100 to +99999 Scaling end of a math channel.
4 Configuration parameters Parameter Value/Selection Description Input signal Configuration ➔ Counter/integrator ➔ Counter/int.chans. ➔ Channel 1 — 6 ➔ Input signal Analog inp. 1 — 3(6), Math channel 1 — 6 The parameter can only be programmed if “Integrator” has been selected under Function. Please select which input is to be integrated. Weighting Configuration ➔ Counter/integrator ➔ Counter/int.chans. ➔ Channel 1 — 6 ➔ Weighting -99999 to +1 to +99999 Select the weighting factor here.
4 Configuration parameters Parameter Value/Selection Description Unit Configuration ➔ Counter/integrator ➔ Counter/int.chans. ➔ Channel 1 — 6 ➔ Unit 5 characters Enter the unit which is displayed on the recorder screen, together with the entered text and the current count. Decimal place Configuration ➔ Counter/integrator ➔ Counter/int.chans. ➔ Channel 1 — 6 ➔ Decimal place Automatic, X.XXXX, XX.XXX, XXX.XX, XXXX.X, XXXXX. Specify the decimal place for displaying the count on the recorder screen.
4 Configuration parameters Parameter Value/Selection Description Ext. Ctrl signal Configuration ➔ Counter/integrator ➔ Ext. ctrl signal Bin. inp. 1 — 4, Logic channel 1 — 6, Low alarm 1 — 6, Low comb.al., High alarm 1 — 6, High comb.al., Count/Int/Al 1 — 6, C/i. comb.al., Comb. alarm, CF inserted, IntMemAlmCF, IntMemAlmSer, MemAlmCFcard, Error, Modbus flag, Only programmable, if the parameter Type is set to “external” on at least one counter/integrator.
4 Configuration parameters 4.2.6 Configuration - Recording Parameter Value/Selection Description Analog channels Input signal Analog input 1 — 6, Configuration Math channel 1 — 6 ➔ Recording ➔ Analog channels ➔ Analog channel 1 — 6 ➔ Input signal Select which channel is to be recorded (stored).
4 Configuration parameters Parameter Differential Value/Selection Description -99999 to 0 to +99999 Configuration ➔ Recording ➔ Analog channels ➔ Analog channel 1 — 6 ➔ Alarm ➔ Differential (1) = Low limit (2) = High limit (3) = Differential Text Low alarm Standard text, Configuration Text 1 — 18, ➔ Recording No text ➔ Analog channels ➔ Analog channel 1 — 6 ➔ Alarm ➔ Text Low alarm Text High alarm Standard text, Configuration Text 1 — 18, ➔ Recording No text ➔ Analog channels ➔ Analog channel 1 — 6 ➔
4 Configuration parameters Parameter Value/Selection Description Event traces Input signal Trace designation Configuration ➔ Recording ➔ Event traces ➔ Event traces 1 — 6 ➔ Input signal Configuration ➔ Recording ➔ Event traces ➔ Event traces 1 — 4 ➔ Trace design. Off, Bin. inp. 1 — 4, Logic channel 1 — 6, Low alarm 1 — 6, Low comb.al., High alarm 1 — 6, High comb.al, Count/Int/Al 1 — 6, C/i. comb.al., Comb.
4 Configuration parameters Parameter Value/Selection Description Storage cycle: Timed operation Configuration ➔ Recording ➔ Timed operation ➔ Storage cycle 1 — 5 — 32767s v Chapter 2.7 “Operating modes” Chapter 2.8 “Storing data” Control signal: Event operation Configuration ➔ Recording ➔ Event operation ➔ Control signal Off, Bin. inp. 1 — 4, Logic channel 1 — 6, Low alarm 1 — 6, Low comb.al., High alarm 1 — 6, High comb.al., Count/Int/Al 1 — 6, C/i. comb.al., Comb.
4 Configuration parameters 4.2.8 Configuration - Control functions Parameter Value/Selection Description External texts: Binary input 1 — 4 Configuration ➔ Control functions ➔ External texts ➔ Binary input 1…4 Standard text, Text 1 — 18, No text v Chapter 3.5 “Event list” External texts: Logic channel 1—6 Configuration ➔ Control functions ➔ External texts ➔ Logic channel 1—6 Standard text, Text 1 — 18, No text Can only be used if the “Math/logic module” option is available. v Chapter 3.
4 Configuration parameters 4.2.
4 Configuration parameters 4.2.11Configuration - Undocumented parameters Parameter Undocumented parameter 1 — 40 Value/Selection Configuration ➔ Undoc. parameter ➔ Par. 1 — 40 Description These parameters are intended for customerspecific adaptation. Alteration to the settings may only be carried on the express instruction of the equipment manufacturer. 4.2.
5 Setup program The setup program is used for the easy creation of configuration files, and to configure the instruments from a PC. 5.1 Hardware and software requirements The following hardware and software requirements have to be met for installing and operating the setup program: Minimum configuration - Intel Pentium1 III or higher, - Microsoft Windows2 NT4.
5 Setup program 5.2 Installation Run the installation program h Start Microsoft Windows H If Microsoft Windows has already been started, all Windows programs must be shut down before installing the setup program. h Insert the CD into the drive, then close the drive. After the CD has been inserted, the installation program will start automatically. If not, proceed as follows: h Start the file “Launch.exe” in the main directory of the CD.
5 Setup program 5.3 User interface Menu bar Toolbar Working area Connection status Menu bar Teleservice The individual functions of the setup program can be started by using the menu bar. v Chapter 5.8 “Menu functions” Toolbar The toolbar contains selected functions of the menu bar. They can be started from the left mouse button. If you rest the mouse pointer on one of the icons (tool tips), you will see the function title after a short delay.
5 Setup program Shifting the toolbar The position of the toolbar can be changed, if desired. h Move the mouse pointer between two icon groups. h Press the left mouse button. h Keeping the left mouse button pressed, drag the toolbar to the required position. h Now release the mouse button.
5 Setup program 5.4 Configuration By using the function File ➔ New (or File ➔ Open) you can create a new configuration file (setup) or open an already existing one. The working area will be filled with the corresponding settings. Current setup Navigation tree for finding the settings quickly Navigation tree Dialog window The settings are displayed here. A single click with the left mouse button in the navigation tree will position the entry visibly in the dialog window.
5 Setup program ... the window becomes an active window. Functions of the right mouse button If you use the right mouse button in the dialog window, different functions will be available. These functions concern that part of the configuration on which the mouse pointer was positioned when the right mouse button was pressed. Example: The right mouse button was pressed when it was on the entry “Analog inputs”. Analog inputs Edit This function starts the change dialog for configuring the analog inputs.
5 Setup program Analog inputs Online edit The current settings of the analog inputs are read out directly from the instrument, and the change dialog is opened. Now you can alter the configuration. After alteration, the new setting is transmitted back to the instrument and entered in the dialog window. If necessary, additional information that is required will be read out from the instrument and also entered in the dialog window.
5 Setup program 5.5 Teleservice You can use Teleservice to request the latest data from a paperless recorder. In addition, the Modbus flag can be switched (activated). In order to use Teleservice, there must be an existing connection to a device. v Chapter 5.7 “Connection between the PC and the recorder” You can switch Teleservice into or out of display via the Window menu.
5 Setup program Setting the Modbus flag h With the left mouse button, click on the arrow pointing to the right ( ), until the register “Additional binary signals” appears on the screen ( ). h Click on the register “Additional binary signals”. This automatically turns it into an active register. h Clicking into the box with the left mouse button (single click) will switch the Modbus flag. Shifting the Teleservice window The Teleservice window, too, can be shifted.
5 Setup program 5.6 Data transfer from and to the instrument There are two ways of transferring the setup data to or from a recorder: - transfer via the CompactFlash memory card and - transfer via interface. H A new configuration will delete the data in the instrument (FLASH memory). If you want to retain the data, then the function CompactFlash card ➔ Update CF card must be started, or the data must be read out with the PCA communications software (PCC) before the configuration is transmitted. 5.6.
5 Setup program Paperless recorder You can use the paperless recorder to write setup data to the CompactFlash memory card, or to read from it. v Chapter 3.6 “CompactFlash card” Import/export errors If an error occurs on the paperless recorder side during transfer from or to a CF card, it will be indicated in the CompactFlash card menu and will remain until the error has been eliminated or is overwritten by a fresh error message.
5 Setup program Formatting the CF card A A CompactFlash memory card (CF card) must be formatted for FAT (not FAT32 or NTFS). If you format a CF card yourself, you must not use fast-formatting (Quickformat). If this is disregarded, a fault-free exchange of data cannot be ensured. To format a CF card using the Windows workplace (e. g. Windows XP): h Make a double-click on the workplace symbol in the “Desktop” area for the PC, using the left mouse button.
5 Setup program 5.6.2 Transfer via interface Data transfer is possible via one of the following interfaces: Paperless recorder PC Setup interface RS232 Setup interface USB RS232 RS232 RS485 RS485 (plug-in card or converter) Ethernet Ethernet v The settings for the interface parameters can be found in Chapter 4.2.10 “Configuration - Interfaces”. Setup - RS232 You will need the interface cable (including adapter) in this case. It is available as an accessory for the recorder.
5 Setup program Ethernet The paperless recorder or a PC can be connected to the network using normal, commercially available network cables (with RJ45 connector). If the recorder and a PC are to be connected directly together, please use a crossover cable. H Starting the transfer Only one 1 PC (client) at a time can access the instrument (server) via the Ethernet interface.
5 Setup program 5.7 Connection between the PC and the recorder If data are to be exchanged between a recorder and a PC, it is essential to define which instrument and which route is used for communication, and to establish a connection between the instrument and the PC. Connection to an instrument is made by using the function Data transfer ➔ Establish connection or by a click on the symbol.
5 Setup program h First select the device version. h Enter an additional description, if appropriate. h Set one of the option fields, if appropriate. h Activate the button. h Select the interface which you want to use to access the device. The next steps depend on the interface or type of connection that has been selected.
5 Setup program TCP/IP PORT Analog modem / ISDN Serial interface The following parameters must be selected: IP address / HOST name xxx.xxx.xxx.xxx Enter the IP address of (Example: 192.168.0.10) your device. If you enter the name, the IP address can be determined by clicking on the button “Convert HOST name to IP address”. Port number, port name 502 The port used for communication. Communications protocol Modbus TCP/IP, Modbus protocol Set the Modbus TCP/IP here.
5 Setup program Transmission rate 9600, 19200, 38400 The transmission rate must match the one set in the device. If “RS232 setup interface (TTL)” has been selected as the control signal, then 9600 must be set. Control signal RS232 If the RS232 interface on the device is used. RS232 setup interface (TTL) If the setup interface on the device is used. RS422-RTS If the RS422/485 interface on the device is used.
5 Setup program Device list When all settings have been made, the device is entered in the device list. 5.7.2 Device list All devices that have been defined are shown in a device list. The interface parameters are also administered here, and new, additional devices can also be defined in the device list. Make a connection with the selected device. Remove device(s) from the list.
5 Setup program 5.8 Menu functions 5.8.1 File New Opens a new setup in the working area. The values will be preset to the factory default settings. Open Opens an existing setup from a file, and present the contents in the working area. Save Saves the setup that is shown in the working area to a file. It is only necessary to enter the file name once. If the file is saved again, no query is made about the file name. Save as Saves the setup that is shown in the working area to a file.
5 Setup program 5.8.2 Edit Undo ... Undoes the last editing action. In the menu, the item Undo shows which setting will be undone. Restore ... The Restore function will only be available when the Undo function has been activated. This function repeats the setting that was previously canceled by the Undo function. Parameterization This function has the same effect as a double-click with the left mouse button on the following function in the dialog window.
5 Setup program Setup data info This function has the same effect as a double-click with the left mouse button on one of the following functions in the dialog window. 5.8.3 Data transfer Make connection This function establishes a connection to a device. A connection to a device is a precondition for transferring a setup to or from a device via an interface (serial or Ethernet). Disconnect This breaks an existing connection.
5 Setup program Enable extra codes This function is intended for later extensions within the paperless recorder (e. g. enabling the extra code for “Integrators, counters, math and logic module”). After starting up the function, the function must be used to read out a code number from the device, and then passed on to the manufacturer. The manufacturer will then generate a “release number”. The function is used to transmit this release number to the device, which will then enable the new device functions.
5 Setup program Date and time This function is used to match the date and time for a PC and a device. Terminates the function. Send the date and time to a device. The button is used to make date and time settings for a device. This function uses the information in the “New setting” selection as the basis for adjustment. ensures that the device clock is read out continually (cyclically). The continual read-out can be ended by activating .
5 Setup program Renew log-in / Alter password The function Renew log-in / Alter password is used to - activate the user and password query at the program start and - alter the current password. This function only has an affect on the operation of the setup program, not the recorder. Activation of the user and password query at the program start When the setup program is installed for the first time, there will not yet be a query of the user name and password when the program starts.
5 Setup program h Enter the password - the “Old password” field remains empty. When the entry has been concluded, the new password is activated by clicking . From now on, the user name and the password will be requested at the program start. H Initially, no start password is assigned to the “Maintenance” user either. At program start, log in with the “Maintenance” user name and enter a password as described above. Password alteration Altering a password is very much like activating the password query.
5 Setup program 5.8.5 Window H The usual Windows options are available for the positioning of the dialog windows. Enlarging and minimizing Reducing and maximizing Close A double-click with the left mouse button produces maximum enlargement. Cascade If several dialog windows are open at the same time, this function has the effect that all windows are shown with an offset to one another. You can use the left mouse button to bring a window into the foreground.
5 Setup program 5.8.6 Info Info on setup Here you can find out the version number of the setup program. Please have the version number available if you contact the service hotline. Software documentation This function calls up the available instrument and software documentation in PDF format. Registered license numbers Here you can find out the license number of the setup program. Please have the license number available if you contact the service hotline.
5 Setup program 5.9 Math/logic The math and logic module is available as an extra code. It has to be ordered from the start, or subsequently enabled through the function Extras ➔ Enable extra codes (Page 98). The math/logic module is composed of channels that are not available as hardware but are calculated by the instrument software. Math You can activate the math channels in the setup program, in the “Math / logic” area, or through the menu Edit ➔ Settings through setup only ➔ Math / logic.
5 Setup program Function selection The function is selected here. All other fields can subsequently be edited according to the function. You must edit the input fields “Variable a”, “Variable b” or “Time for moving average” if one of the standard functions (difference, ratio, humidity, moving average) has been set. If you have set the “Formula” function, then the input field “Formula” must be edited. Entry can either be direct or via a ( ) dialog.
5 Setup program Example: Formula The formula can be entered directly from the PC keys or by calling up the function. When you have called up the function, this dialog box will appear: In the left window you can select the required signal, in the right window the required operator, and enter them into the formula by activating the corresponding button ( ). Instead of using , you can also make a selection by a double-click with the left mouse button.
5 Setup program Operator overview Operator priorities Operator + * / ( ) SQRT() MIN() MAX() SIN() COS() TAN() ** EXP() ABS() INT() FRC() LOG() LN() Explanation addition subtraction multiplication division opening bracket closing bracket root minimum value maximum value sine cosine tangent x to the power y exponential function absolute value integer portion decimal portion logarithm natural logarithm Example AE1 + AE2 AE1 - AE2 AE1 * AE2 AE1 / AE2 ( ) SQRT (AE1) MIN (AE1, AE2) MAX (AE1, AE2, AE3) SIN (AE
5 Setup program 5.10 Character set 032 033 034 035 036 037 038 039 040 041 042 043 044 045 046 047 048 049 050 051 052 053 054 055 056 057 058 059 060 061 062 063 064 065 066 067 068 069 070 071 072 073 074 075 076 077 078 079 ! " # $ % & ’ ( ) * + , .
5 Setup program Entering special characters Any (special) characters that cannot be entered directly from the keyboard of the PC can be entered with the help of the A key and the numerical combinations that are specified in the table. Example The special character © has to be entered: h Position the cursor with the mouse, or by using the cursor keys, on the point where the character is to be inserted.
6 Rights 6.1 Rights with regard to the setup program Depending on the installation and the log-in, individual users will have different rights within the setup program.
6 Rights 112
7 PC programs 7.1 PC evaluation software (PCA3000) H Operating Instructions 59498 describe PCA300 in detail. PCA3000 is the professional evaluation software for data-analysis of the archived instrument process data. PCA3000 is the ideal software for the best-quality graphic and alphanumeric display of electronically stored measurement data. The software supports multi-user functionality, so that different users can access the same data. PCA3000 runs under Windows NT, Windows 2000 and Windows XP.
7 PC programs 7.2 PCA communication software (PCC) H PCC is described in detail in the Operating Instructions 59500. PCC is the professional archive software: it transfers the data via an inerface (setup, serial or Ethernet) from the recorder (or several recorders and controllers), and deposits them on a PC or a network.
8 Identifying the instrument version 8.
8 Identifying the instrument version 8.2 Standard accessories - 1 Operating Instructions 59480 - 2 mounting brackets, or 4 mounting brackets with IP65 - 1 control panel seal with IP65 - 2 cover plates for fixing holes - Cable tie with foot (can be released) for strain relief of the sensor connecting cables 8.
9 Installation 9.1 Installation site and climatic conditions The installation site should be free from vibration. Electromagnetic fields, caused by motors, transformers etc. should be avoided. The ambient temperature at the site should be 0 to +45°C, with a relative humidity of 75%, no condensation. v Chapter 10.1 “Installation notes” 9.2 Fitting Views Size 26 increases to 27 if the IP65 seal is used.
9 Installation Installation h Insert the paperless recorder from the front into the panel cut-out. h From behind the panel, hook the two mounting brackets into the grooves on the sides of the housing. The flat faces of the mounting brackets must lie against the housing. Cutout for fixing bracket h Push the fixing brackets up against the back of the panel and tighten them evenly. h Use the cover plates to close the remaining cutouts for mounting brackets.
9 Installation Extra code IP65 The following instructions must be followed when installing the paperless recorder with extra code IP65. - The seal for the control panel must be fitted before the recorder is installed. Caution! Make sure that the seal is correctly positioned. The two large indentations on the front frame must be closed by the two lugs on the seal. - Instead of the 2 mounting brackets, there are 4 wide mounting brackets that have to be used (one element for each side of the housing).
9 Installation 120
10 Electrical connection 10.1 Installation notes k The choice of cable, the installation and the electrical connection must conform to the requirements of VDE 0100 “Regulations on the Installation of Power Circuits with Nominal Voltages below 1000V” or the appropriate local regulations. k Work inside the instrument must only be carried out to the extent described and, like the electrical connection, only by qualified personnel.
10 Electrical connection 10.3 Connection diagram V The electrical connection must only be carried out by qualified personnel.
10 Electrical connection Terminal assignment (for pluggable screw terminals) 3/6 channel paperless recorder Analog inputs Connector Thermocouple 1 to 6 Diagram If sizeable supply interference is present on a thermocouple input, then the corresponding analog input requires a link from terminal 2 to terminal 4. Increased interference voltages can, for example, be generated by ceramic insulation becoming conductive at temperatures of several hundred °C.
10 Electrical connection Terminal assignment Diagram Setup interface (included in delivery) The setup interface can be found on the front of the instrument, behind the cover for the CompactFlash memory card.
11 Index A Action (outputs) 73 Active operating mode 23 Alarm 70 Alarm (counter/integrator) 68 Alarm delay 71 Alarm text (counter/integrator) 68 Alter password 101 Analog inputs 64–65, 70–71 Arrange symbols 103 Authorization code 64 B Basic menu 11, 32 Battery empty 53–54 Baud rate 75 C Calibration status 76 Cascade 103 CF card code number 63 Channel (counter/integrator) 66 Channel indication 35, 61 Channel line 14, 61, 70 Channel name 70 Channel name (counter/integrator) 67 Channel representation 14, 33,
11 Index Connection circuit 64 Connection diagram 122 Connection status 80, 103 Contrast 60 Control function 74 Control signal 60, 73 Control signal (outputs) 73 Counter 66 Counter frequency (counters/integrators) 17 Counter text (counter/integrator) 68 Counters 17 Create screen printout 100 D Daily end time (counter/integrator) 69 Daily start time (counter/integrator) 69 Data Evaluation 27 Read-out via interface 27 Data export to CF card 98 Data format 75 Data import from CF card 98 Data lost 53–54 Data t
11 Index Authorization code 64 Computed code No. 64 Enable options 64 Enable program options 98 End temperature 65 End time (timed operation) 72 End value Target 76 Enter key 11 Error 54 Ethernet 28, 75 Evaluation of measurement data 38 Evaluation program 50 Event definition 47 Event list 47 Event operation 23 Storage cycle 73 Stored value 73 Event traces 15 Exit 96 Exit key 11 Ext.
11 Index Input signal (counter/integrator) 66–67 Installation 117 Installation notes 121 Installation site 117 Integrator 17, 66 Interface type 75 Internal analog inputs 14 IP address 75 IP65 119 K Key inhibit 74 Keys 10, 12 L Language 63 Life-cycle data management 24 Limit index 35 Limit value (counter/integrator) 68 Line width 70 Linearization 64 Logic module 22, 105 Operator priorities 108 Low limit 70 M Make connection 98 Math module 20, 105 Operator overview 108 Operator priorities 108 Meas.
11 Index O Open 96 Operating modes 23 Operating time counter 18 Outputs 16, 73–74 Action 73 Control signal 73 Overrange 14 P Panel mounting 117 Paper perforations 61 Parameterization 60 Password 45–46, 50, 57, 59, 63 Password activation 101 Password alteration 101–102 PCA3000 20, 27, 64, 113, 116 PCC 27, 86, 91, 113–114, 116 Peak value 40 Period (counter/integrator) 68 Power LED 11 Principle of operation 46 Print 96 Print preview 96 Printer setup 96 Program folder 104 Program options, enabling 98 Protocol
11 Index Scaling end 65–66 Scaling start 65–66 Screen off 44 Screen saver 11, 44, 60 Screen texts 10 Scroll operation 38 Sensor 64 Sensor monitoring 14 Setup program Assistant for device settings 91 CompactFlash 86 Configuration 81 Current setup 81 Data transfer 86 Device list 95 Hardware and software requirements 77 Installation 78 Menu bar 79 Menu functions 96 Modbus flag 84–85 Navigation tree 81 Program start 78 Shifting the Teleservice window 85 Teleservice 80, 84 Toolbar 79 Toolbar shifting 80 Transfer
11 Index T Teleservice 103 Temperature unit 63 Text entry 55 Text High alarm 71 Text Low alarm 71 Texts 47, 74 Threshold value (counter/integrator) 67 Tiled horizontally 103 Time 62 Time base (counter/integrator) 67 Time synchronization 62 Timed operation 23 End time 72 Start time 72 Trace design.
11 Index 132
2008-11-20/00578800 Austria China France Germany UK USA T: +43 (0) 2236 691 121 T: +86 22 8398 8098 T: +33 (1) 77 80 90 40 T: +49 (0) 561 505 1307 T: +44 (0) 1273 606 271 T: +1 800 866 6659 Email Website Inquiries@West-CS.com www.West-CS.