Specifications
Alto 818 Installation and Operation
8-4 P/N 66109074001 Rev. D
8.3 TERMINAL SET-UP
The following procedure for setting terminal communication parameters applies to the Qume
QVT 100+ terminal. However, almost any ANSI terminal can be used to configure the Alto 818
(refer to the individual manufacturers' user manuals for terminal setup procedures).
The Alto 818 is shipped with the terminal port set at 300 baud. The baud rate may be changed
to 1200, 2400, 4800, or 9600 as appropriate. A terminal used with the Alto 818 should be set for
odd parity, 7 data bits, and 1 stop bit.
The Qume QVT 100+ (or equivalent) terminal is set up using the following procedures.
1. Simultaneously press SHIFT and SETUP keys. Setup menu displays at foot of screen. Scroll
down menus to SET 4 EIA=HOST using down arrow key. Scroll across this menu using right
arrow key. Using space bar to alter settings, set parameters to read as follows:
SET 4|EIA=HOST|7 BITS|BIT 8 1|PARITY ON|PARITY ODD|STOP 1|BAUD:
300
After all settings are correct, press SETUP key to exit the setup mode.
2. Locate and press the reset switch (S1) (see Figure 2-3) on the ACU; a sample screen display
follows:
MON 05-JUN-95
The date display is followed by log 01 giving the time and the reset message, log 02 with the
tamper message, and log 03 with the power fail message. For example:
01 02:24:44 RESET
02 02:24:50 TAMPER
03 02:24:50 POWER FAIL
If any system faults are detected, log 03 will be followed by further data log entries which report
failures. Each failure report will be preceded by a number that denotes the sensor port from which
the fault was detected. For example:
04 02:24:55 (4) MSM FAIL
05 02:24:57 (1) VIP TAMPER
Log 04 indicates a failure of the MSM connected at sensor port 4; log 05 indicates the tamper
switch on door 1 VIP is active. If no system faults are indicated, the system is ready to log on
and set up the database.