IN_IP5_Manual8.qxd 11/28/06 1:12 PM Page 1 Using the iP5 to Program Inteo Selection of Inteo Instrument Style IN-CIC Custom CIC Fits up to 65-70 dB Hearing Loss IN-X Custom ITE or ITC with High Definition Locator Fits up to 70-75 dB Hearing Loss IN-9 BTE with High Definition Locator Fits up to 70-75 dB Hearing Loss IN-9é Open Fit BTE with High Definition Locator Fits up to 25-30 dB Hearing Loss below 500 Hz and 70-75 dB above 1000 Hz.
IN_IP5_Manual8.qxd 11/28/06 1:12 PM Page 2 iP5 is now in the Pre-Conditions mode. B 1. Choose [M-B] in the mode window (See B) to designate fitting as monaural or binaural. Use [ARROW] key to choose MON or BIN. Press [BACK] to exit [M-B]. An exclamation point (!) will appear next to [M-B] in the mode window to indicate that a monaural fitting has been chosen. 2. Select [VNT] to choose the vent diameter in millimeters (mm). Use the up or down [ARROW] key for selection. Press [BACK] to exit [VNT].
IN_IP5_Manual8.qxd 11/28/06 1:12 PM Page 4 Fine Tuning Inteo B 1. Turn on iP5: Press [on R] for right or [on L] for left (See A). Note: Only one ear can be fine tuned at a time. A 2. Press [MODE] if iP5 is not in FINE TUNE mode (See B). When in FINE TUNE mode you may access the Loudness Master [LNS], Insertion Gain [IG], Maximum Power Output [M P O] including Automatic Output Control [AOC] and Occlusion Manager [O C C]. 1. B To access Loudness Master press [LNS] (see B).
IN_IP5_Manual8.qxd 11/28/06 1:12 PM Page 6 • To access Automatic Output Control press [= =>] in the FINE TUNE mode followed by [MPO]. • Press [AOC] (see B). • Press [UP ARROW] beneath [OFF]. • Press [BACK] (see B) to return to [MPO]. An exclamation point (!) will appear in the mode window next to [AOC] to indicate that a change has been made to Automatic Output Control. Note: When Automatic Output Control is turned off MPO values return to target and further changes cannot be made. 4.
IN_IP5_Manual8.qxd 11/28/06 1:12 PM Page 8 3. Overall gain may be adjusted, in one dB steps for each acoustic listening program other than the primary Master Program and the Audibility Extender, in the [P ROGRAM GAIN OFFSET] mode. B e) Press [BAC K ] (see B) to return to [PROGRAM OPTIONS]. An exclamation point (!) will appear in the mode window next to [PRG] to indicate the selection of additional listening programs. Note: [PGO] can only be accessed if additional listening programs have been selected.
IN_IP5_Manual8.qxd 11/28/06 1:12 PM Page 10 5. The Speech and Noise mode settings can be adjusted for each acoustic listening program, except Audibility Extender, in the [SPEECH – NOISE] mode. Note: Audibility Extender is displayed as [* AE *] in this screen to indicate that the listening program has been selected but no change to the speech and noise setting can be made. B f) Press [BACK] (see B) to return to [P ROGRAM OPTIONS].
IN_IP5_Manual8.qxd 11/28/06 1:12 PM Page 12 Inteo Test Mode The Test Mode allows the clinician to compare the output of the hearing aid against factory specifications. The hearing aid is set to a fixed frequency gain setting during this measurement. B 1. Turn off the iP5 by pressing [ON R] AND [ON L] simultaneously. 2. Connect the Inteo hearing aid to the appropriate inlet [ON L or ON R]. 3. Hold down the top right menu key while pressing the appropriate [ON] key. 4. [TEST MODE] is displayed.