Specifications
InteoProgram Options
When in  the  PROGRAM OPTIONS mode  you may access Program Selection [PRG], Remote
Control matching [RC3], Program Gain Offset [PGO], High Definition Locator [LOC], Speech
and Noise [SN] and Feedback Cancelling [FBC].
1. The default setting for programs [PRG] is the Master Program. As many as two acoustic and
two telecoil programs may be selected in addition to the basic Inteo Master program for a
total  of  five.  The  Inteo  CIC may  have  a  total  of  three  acoustic  programs  which  can  be
accessed only when using a remote control. The Inteo Master Program provides the best
overall performance in all environments. Each of the other seven available program options
has a special focus. Those programs are:
• [T] telecoil • [TV] television
• [MT] microphone plus telecoil • [MUS] music
• [AE] audibility extender • [ACC] acclimatization
• [COMF] comfort
The  dispenser has  the flexibility  of placing  programs  2  through  5  in  their  preferred  order.
The Master program is always in the first program position when programming with the iP5 
programmer.
a) Press  [MODE]  (see  B)  to  access  [PRO-
GRAM OPTIONS]. 
b) When  viewing  the  PROGRAM OPTIONS
screen note that [P1] beside MAST flashes
to  indicate  that  the  first  program  is  the
Master  Program.  Listening  programs that
h a ve  previously  been  selected  will  be 
d i s p l ayed   in  the  P ROGRAM  OPTIONS
window but cannot be changed there.
c) Press [PRG] in the mode window (see B). [PROGRAM SELECTION] will be displayed.
[P1] beside MAST flashes.
d) The second program [P2] position is indicated in the lower left of the [PROGRAM SELEC-
TION] screen by a dash [-]. A listening program may be added to this position by press-
ing  the  up  or  down  [ARROW  KEY]  beneath to  scroll through  the  available  listening 
program selections. The order of the program selections, when using the [UP ARROW] to
scroll, is as follows:
• Master [MAST] • Audibility Extender [AE]
• Acclimatization [ACC] • Microphone plus Telecoil [MT]
• Music [MUS] • Telecoil [T]
• Television [TV] • No program [-]
• Comfort [COMF]
The order is reversed when scrolling with the [DOWN ARROW] key.
Use the same procedure to add or change a listening program selection in each available
program position.
• To access Automatic Output Control press [= =>] in the FINE  TUNE mode followed 
by [MPO].
• Press [AOC] (see B).
• Press [UP ARROW] beneath [OFF].
• Press [BACK] (see B) to return to [MPO]. An exclamation point (!) will appear in the
mode window  next to [AOC]  to  indicate that  a  change has  been made  to  Automatic
Output Control.
Note: When Automatic Output Control is turned off MPO values return to target and further
changes cannot be made.
4. The  default  setting  for  Inteo  Occlusion
Manager  is  [O C C ]  off.  The  Occlusion
Manager  can  be  turned  o n m a n u a l l y.
Turning  Occlusion Manager  on de-couples
the low fre q u e n c y channels, below 500 Hz,
f r om  the higher fre q u e n cy channels and
i n c rea ses compression for the very low fre-
q u e n cy channels. It further allows for man-
ual gain adjustment of the isolated low fre-
q u e n cy re g i o n s . Ty p i c a l l y, gain is decre a s e d
by  making  negative  offsets  for  LF1  and/ 
or LF2.
• To access Occlusion Manager press [= =›] in the Fine Tune mode (see B) followed by
[OCC]. An asterisk (*) appears above [OFF ].
• Press [UP ARROW] beneath [ON]. An asterisk (*) now appears above [ON]. To compare
the two settings alternate between [ON] and [OFF] by pressing the [ARROW KEYS] and
ask the client to select the one that makes their voice sound better. Ask the client to say 
“Baby Jeanie is teeny tiny” and note the effect on the hollowness of their voice as they
repeat the phrase.
• For  further  fine  tuning  of  the  Occlusion  Manager  changes  may  be  made  to  LF1
(LF1==125,250 and 350 Hz) or LF2 (LF2= 350 and 500 Hz). Gain in the desired chan-
nel can be decreased by pressing the [DOWN ARROW]. 
Note: If the occlusion is believed to be related to too much low frequency gain then decreasing
LF1 will help to alleviate the problem. Change LF1 for “loudness issues; change LF2 for “clar-
ity” adjustment.
• LF1 is not available on the Inteo élan. 
• Press [BACK]( see B) to return to FINE TUNE. An exclamation point (!) will appear in the
mode window next to [OCC] to indicate that a change has been made to the Occlusion
Manager.
Note: If the occlusion is related to the actual physical fit of the hearing aid activating the 
occlusion manager may NOT resolve the difficulty.
7
6
B
B
IN_IP5_Manual8.qxd 11/28/06 1:12 PM Page 6