Differential pressure transmitter DPT10 - metal measuring diaphragm Profibus PA Document ID: 37245
Contents Contents 1 About this document 1.1 1.2 1.3 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . .. .. .. .. .. .. .. .. .. 5 5 5 5 6 6 6 6 6 . . . . . . . . . . . . . . . . . . . . . . . . .. .. .. .. 7 8 12 12 General instructions . . . . . . . . . . . . . . . . . . . . . . . Special application conditions . . . . . . . . . . . . . . . . Measurement setup flow . . . . . . . . . . . . . . . . . . . . Measurement setup level . . . . . . . . . . . . . . . . . . .
Contents 6.6 6.7 7 Parameter adjustment with PDM . . . . . . . . . . . . . . . . 59 Set up 8.1 8.2 8.3 8.4 8.5 9 55 58 Setup with PDM 7.1 8 Menu schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . Saving the parameter adjustment data . . . . . . . . . . . . .. .. .. .. .. 60 60 62 66 66 Maintain. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Rectify malfunctions . . . . . . . . . . . . . . . . . . . . . . . . . Instrument repair . . . . . . . . . . . . . . . . .
1 About this document 1 About this document 1.1 Function This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instructions for maintenance and fault rectification. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device. 1.2 Target group This operating instructions manual is directed to trained qualified personnel.
2 For your safety 2 For your safety 2.1 Authorised personnel Mount and set up the pressure transmitter only if you know the applicable national regulations and have the appropriate qualification. You must be aquainted with the regulations and instructions for hazardous areas, measurement and control technology as well as electrical circuits because the pressure transmitter is "electrical equipment" according to EN 50178.
2 For your safety 2.5 Safety label on the instrument The safety approval markings and safety tips on the device must be observed. 2.6 CE conformity This device fulfills the legal requirements of the applicable EC guidelines. By attaching the CE mark, we provide confirmation of successful testing. 2.7 Fulfillment of NAMUR recommendations The device fulfills the requirements of the applicable NAMUR recommendations. 2.
3 Product description 3 Product description 3.
3 Product description l l Hardware and software version Article numbers, documentation 3.2 Principle of operation Application area DPT10 is a differential pressure transmitter for measurement of flow, level, differential pressure, density and interface. Measured products are gases, vapours and liquids. Flow measurement + Q ~ ∆p p1 + – p2 Q Q ~ ∆p – p1 p2 Q 1 2 Fig.
3 Product description Level measurement – ∆p h= ρ g h – + + – + 2 1 3 Fig. 3: Level measurement with DPT10. Δp = differential pressure, ρ = density of the medium, g = acceleration of gravity 1 2 3 Basic version with effective pressure lines Version with flange isolating diaphragm Version with capillaries and cell isolating diaphragms Differential pressure measurement 2 + 1 Fig.
3 Product description Density measurement – = ∆p h g h + Fig. 5: Density measurement with DPT10, h = defined mounting distance, Δp = differential pressure, ρ = density of the medium, g = acceleration of gravity 1 DPT10 Interface measurement – 2 + h 3 1 Fig. 6: Interface measurement with DPT10 1 2 3 Functional principle DPT10 Liquid with highest density Liquid with lowest density A metallic measuring cell is used as sensor element.
3 Product description The configuration of the measuring cells differs depending on the measuring range: 1 2 p1 3 4 p2 5 Fig. 7: Metallic measuring cell 10 mbar and 30 mbar - p1 and p2 process pressures 1 2 3 4 5 Measuring element Silicone diaphragm Separating diaphragm Filling oil Integrated overvoltage arrester 1 p1 2 3 4 p2 Fig.
3 Product description a PDM environment, an EDD (Electronic Device Description) is also required to enable the full range of sensor functions (also available as a download). A CD with the appropriate files can be ordered via e-mail or by phone from one of the WIKA agencies. The backlight of the indicating and adjustment module is powered by the sensor. Prerequisite is a certain level of operating voltage. The data for power supply are specified in chapter "Technical data". 3.
3 Product description Storage and transport temperature l l l Not exposed to corrosive media Protected against solar radiation Avoiding mechanical shock and vibration l Storage and transport temperature see chapter "Supplement Technical data - Ambient conditions" Relative humidity 20 … 85 % 37245-EN-100120 l Differential pressure transmitter DPT10 - metal measuring diaphragm • Profibus PA 13
4 Mounting 4 Mounting 4.1 General instructions Suitability for the process conditions Make sure that all parts of the instrument exposed to the process, in particular the sensor element, process seal and process fitting, are suitable for the existing process conditions. These include above all the process pressure, process temperature as well as the chemical properties of the medium. You can find the specifications in chapter "Technical data" or on the type label.
4 Mounting Effective pressure lines You find general recommendations for wiring of effective pressure lines in DIN 19210 "Effective pressure lines for flow systems" or the corresponding national or international standards. When wiring effective pressure lines outdoors, keep in mind to use a suitable antifreeze, e.g. by using tube heat tracing. Wire effective pressure lines with a monotonic decrease of at least 10 %.
4 Mounting 4.3 Measurement setup flow In gases à Mount DPT10 above the measurement loop so that condensate can drain off in the process cable. 1 2 3 + – 4 Fig.
4 Mounting 1 2 3 3 + 4 – 5 5 6 6 7 Fig. 11: Measurement setup, flow measurement in vapours 1 2 3 4 5 6 7 In liquids Condensate vessels Orifice or impact pressure probe Blocking valves DPT10 Precipitator Drain valves 3-fold valve block à Mount DPT10 below the measurement loop so that the effective pressure lines are always filled with liquid and gas bubbles can bubble up to the process line à For measurements in products with solid content such as e.g.
4 Mounting 1 2 + 2 3 – 4 4 5 5 6 Fig. 12: Measurement setup, flow measurement in liquids 1 2 3 4 5 6 Orifice or impact pressure probe Blocking valves DPT10 Precipitator Drain valves 3-fold valve block 4.
4 Mounting à For measurements in products with solid content such as e.g. dirty liquids, the installation of separators and drain valves is recommended to enable collection and removal of debris and sediment. patm + min. patm 1 2 4 5 3 Fig.
4 Mounting In closed vessels with effective pressure lines à Mount DPT10 below the lower measurement connection so that the effective pressure lines are always filled with liquid à Connect minus side always above the max. level à For measurements in products with solid content such as e.g. dirty liquids, the installation of separators and drain valves is recommended to enable collection and removal of debris and sediment. – 1 max. min. + 1 2 3 3 4 4 5 Fig.
4 Mounting 1 – max. min. 2 + – + 3 4 Fig. 16: Measurement setup, level measurement in closed vessel 1 2 3 4 In closed vessels with bilateral pressure transmitter Blocking valve Precipitator Drain valve DPT10 à Mount DPT10 below the lower isolating diaphragm à The ambient temperature should be the same for both capillaries Information: Level measurement is only ensured between the upper edge of the lower and the lower edge of the upper pressure transmitter. – max. 1 min.
4 Mounting à Connect minus side always above the max. level à The condensate vessel ensures a constant pressure on the minus side à For measurements in products with solid content such as e.g. dirty liquids, the installation of separators and drain valves is recommended to enable collection and removal of debris and sediment. – 1 max. 2 min. + 2 3 4 5 5 6 Fig.
4 Mounting – 1 max. 2 min. 3 + – + 5 4 Fig. 19: Measurement setup in closed vessel with superimposed steam 1 2 3 4 5 Condensate vessel Blocking valve Precipitator Drain valve DPT10 4.5 Measurement setup density and interface Density and interface measurement are carried out in level measurement mode.
4 Mounting Interface measurement à Mount DPT10 below the lower isolating diaphragm à The ambient temperature should be the same for both capillaries Information: An interface measurement is only possible if the densities of the two media remain the same and the interface is between the two measurement points. The total level must be above the upper measurement point. 0,3 m 0,8 1,0 – + 1 Fig. 21: Measurement setup with interface measurement. Min. adjustment with filling complete with density 0.
4 Mounting 4.6 Measurement setup differential pressure In gases and vapours à Mount DPT10 above the measurement loop so that condensate can drain off in the process cable. 1 2 + 3 3 4 Fig. 22: Measurement setup, differential pressure measurement in gases and vapours 1 2 3 4 In liquids DPT10 3-fold valve block Blocking valves E.g.
4 Mounting 1 2 2 – + 3 4 4 5 5 6 Fig. 23: Measurement setup, flow measurement in liquids 1 2 3 4 5 6 In gases, vapours and liquids E.g. filter Blocking valves DPT10 Precipitator Drain valves 3-fold valve block à Mount isolating diaphragm with capillaries on top or laterally on the pipeline à In vacuum applications: Mount DPT10 below the measurement loop à The ambient temperature should be the same for both capillaries 2 1 2 – + 3 4 1 2 3 4 26 Isolating diaphragm with bolting Capillaries E.
4 Mounting 4.7 Mounting arrangement and connection valve block Mounting arrangement The following illustration shows the elements for a tube mounting and an example for a mounting arrangement with valve block. 1 2 3 4 5 + – 6 7 8 9 Fig.
4 Mounting Connection valve block The following illustration shows the connection of a 5-fold valve block. A 7 8 9 A: 3 5 1 4 2 3 6 4 5 1 7 9 8 2 6 Fig.
4 Mounting 8 9 Oval flange adapter Fixing screws 4.8 External housing Wall mounting 1 Mark the holes according to the following drilling template 2 Depending on the mounting surface, fasten the wall mounting plate with 4 screws 90 mm (3.54") 70 mm (2.76") 3 mm (0.12") 8 mm (0.32") 93 mm (3.66") 110 mm (4.33") mm ,5 ") R3 0.14 ( Fig. 27: Drilling template - wall mounting plate Mount the wall mounting plate so that the cable entry of the socket housing points downward.
5 Connecting to power supply 5 Connecting to power supply 5.
5 Connecting to power supply Select connection cable for Ex applications Take note of the corresponding installation regulations for Ex applications. In particular, make sure that no potential equalisation currents flow over the cable screen. In case of grounding on both sides this can be achieved by the use of a capacitor or a separate potential equalisation. 5.
5 Connecting to power supply The electrical connection is finished. Fig.
5 Connecting to power supply 2 Remove the housing socket from the mounting plate 3 1 2 Fig. 29: Components of the external housing 1 2 3 Screw Wall mounting plate Cable gland 3 Loop the connection cable through the cable entry on the housing base1) Information: The cable gland can be mounted in three positions each displaced by 90°. Simply exchange the cable gland against the blind plug in the suitable thread opening.
5 Connecting to power supply 5.3 Single chamber housing The following illustrations apply to the non-Ex as well as to the Ex-ia version. Electronics and connection compartment Display I²C 1 4 2 5 6 7 8 1 2 3 Fig.
5 Connecting to power supply 5.4 Double chamber housing The following illustration apply to non-Ex as well as Ex ia versions. The Exd version is described in the next subchapter. Electronics compartment 1 Display I²C 1 2 5 6 7 8 3 2 Fig.
5 Connecting to power supply Display Connection compartment 1 3 1 I²C 2 2 Fig. 33: Connection compartment double chamber housing 1 2 3 Plug connector for service Ground terminal for connection of the cable screen Spring-loaded terminals for voltage supply Wiring plan I2C 1 2 1 Fig.
5 Connecting to power supply 5.5 Double chamber housing Ex d Electronics compartment 1 Display I²C 1 2 5 6 7 8 2 3 Fig. 35: Electronics compartment, double chamber housing 1 2 3 Plug connector for service Internal connection cable to the connection compartment Terminals for the external indicating and adjustment module Connection compartment 1 1 2 37245-EN-100120 2 Fig.
5 Connecting to power supply Wiring plan 1 2 1 Fig. 37: Wiring plan with double chamber housing Ex d 1 Voltage supply/Signal output 5.6 Version IP 66/IP 68, 1 bar Wire assignment, connection cable + 1 2 Fig.
5 Connecting to power supply 5.7 External housing with version IP 68 Display Electronics and connection compartment for power supply I²C 5 1 2 5 6 7 1 8 2 4 3 Fig.
5 Connecting to power supply Terminal compartment for sensor connection 1 2 3 4 1 2 3 4 5 Fig. 40: Connection of the sensor in the housing socket 1 2 3 4 5 Brown Blue Yellow White Shielding Wiring plan external electronics Display I2C 1 2 5 6 7 8 1 Fig. 41: Wiring plan external electronics 1 Voltage supply 5.8 Switch on phase 40 After DPT10 is connected to voltage supply or after voltage recurrence, the instrument carries out a self-check for approx. 30 seconds.
5 Connecting to power supply l l l Internal check of the electronics Indication of the instrument type, the firmware as well as the sensor TAGs (sensor designation) Status byte goes briefly to fault value 37245-EN-100120 Then the current measured value will be displayed and the corresponding digital output signal will be outputted to the cable.2) 2) The values correspond to the actual measured level as well as to the settings already carried out, e.g. default setting.
6 Adjustment with the indicating and adjustment module 6 Adjustment with the indicating and adjustment module 6.1 Short description Function/Configuration The indicating and adjustment module is used for measured value display, adjustment and diagnosis.
6 Adjustment with the indicating and adjustment module Fig. 42: Insert indicating and adjustment module 37245-EN-100120 Note: If you intend to retrofit the instrument with an indicating and adjustment module for continuous measured value indication, a higher cover with an inspection glass is required.
6 Adjustment with the indicating and adjustment module 6.3 Adjustment system 2 1 1.1 3 Fig.
6 Adjustment with the indicating and adjustment module 6.4 Parameter description Introduction DPT10 has general adjustment parameters which are also used for other measuring principles as well as instrument-specific adjustment parameters. The general adjustment parameters are described in the operating instructions manual "Indicating and adjustment module". The instrument-specific adjustment parameters are described in this chapter.
6 Adjustment with the indicating and adjustment module Unit of measurement In this menu item you select the adjustment unit as well as the unit for the temperature indication in the display. To select the adjustment unit (in the example switching over from mbar to bar), proceed as follows: 1 ▶ 2 Push the [OK] button in the measured value display, the menu overview is displayed.
6 Adjustment with the indicating and adjustment module Unit of measurement Density 0001000 kg/dm³ 6 Enter the requested density value with [->] and [+], confirm with [OK] and move to position correction with [->]. The adjustment unit is thus switched over from bar to m. Proceed as follows to select the temperature unit: à Activate the selection with [OK] and select "Temperature unit with [->]. à Activate the selection with [OK] and select the requested unit with [->] (e.g. °F). à Confirm with [OK].
6 Adjustment with the indicating and adjustment module If a known value should be taken over as position correction which is not the current value, then you have to select the function "Edit" and enter the requested value. Zero adjustment with differential pressure In this menu item, the min. differential pressure is entered. Proceed as follows: 1 Edit the bar value in the menu item "zero" with [OK]. Zero adjustment 0.00 % = 0.0000 bar 0.0000 bar DP 2 Set the requested value with [+] and [->].
6 Adjustment with the indicating and adjustment module Min. adjustment with level Proceed as follows: 1 Edit the % value in the menu item "Min. adjustment" with [OK]. Min. adjustment 0.00 % = 0.0000 bar 0.0000 bar 2 Set the requested value with [+] and [->]. 3 Confirm with [OK] and edit the requested bar value. 4 Set the requested bar value with [+] and [->]. 5 Confirm with [OK] and move to max. adjustment with [->].
6 Adjustment with the indicating and adjustment module 2 Set the requested bar value with [+] and [->]. 3 Confirm with [+] and move to max. adjustment with [->]. For an adjustment with flow, simply enter the actual measured value indicated at the bottom of the display. Information: The DPT10 is also suitable for bidirectional flow measurement (flow in both directions). The selection is carried out in the menu item "Linearization curve". With the bidirectional flow measurement, the min.
6 Adjustment with the indicating and adjustment module Linearisation curve linear Enter the requested parameter via the appropriate keys, save your settings and jump to the next menu item with the [->] key. Leak flow volume suppression with flow In some application, small flow quantities should not be detected. With the creeping quantity suppression, the flow value can be suppressed up to a certain % value. The default value is 5 % of the max. flow value, corresponding to 0.25 % of the max.
6 Adjustment with the indicating and adjustment module Part sum counter 0 % = +0000 m3/s= 100 % = +0000 m3/s= 6 Edit with [OK] and set the requested values with [+] and [->]. 7 Confirm with [OK] and jump back to the indication of the part sum counter. 8 Select with [->] the unit of the sum counter, adjust the requested unit with [->], for example m3/s and confirm with [OK]. The setting of the part sum counter is hence terminated, the counting function is activated.
6 Adjustment with the indicating and adjustment module Menu item Reset value Basic settings Unit of measurement bar Temperature unit °C Zero/Min. adjustment Measuring range begin Span/Max. adjustment Measuring range end Density 1 kg/l Density unit kg/l Damping 1s Linearisation linear Sensor-TAG Sensor Display Displayed value PA-Out Diagnostics Total sum counter 0.0000 1000 gal Part sum counter 0.
6 Adjustment with the indicating and adjustment module 6.5 Menu schematic Information: Depending on the version and application, the highlighted menu windows are not always available. Basic adjustment differential pressure ▶ Basic adjustment Display Diagnostics Service Info Sensor address 1 1.1 126 Application 1.1 Differential pressure ▼ Zero 000.0 % = 0.0000 bar 0.0000 bar 1.3 Sensor-TAG 1.7 DP Max. adjustment 100.00 % = 0.5000 bar 0.0000 bar 1.
6 Adjustment with the indicating and adjustment module 6.6 Menu schematic Basic adjustment flow ▶ 1 Basic adjustment Display Diagnostics Service Info Sensor address 1.1 126 Min. adjustment 000.0 % = 0.0000 bar 0.0000 bar Application 1.1 Flow ▼ 1.3 - + DP Leak volume suppression 1.6 Activated ▼ Max. adjustment 100.00 % = 0.5000 bar 0.0000 bar 1.4 Sensor-TAG 1.7 - Unit Unit of measurement bar▼ Temperature unit °C▼ 1.1 Damping 1.5 Position correction Offset = -0.0035 bar 0.0000 bar 1.
6 Adjustment with the indicating and adjustment module Display ▶ 2 Basic adjustment Display Diagnostics Service Info 2.1 Displayed value Differential pressure ▼ Lighting Displayed value 2.1 Display unit 2.2 Scaling 0 % = 0.0 l 100 % = 100.0 l 2.3 3.3 Total sum counter 0.0000 1000 gal 3.4 Volume ▼ l▼ PA-Out 2.4 Switched off ▼ Diagnostics ▶ Basic adjustment Display Diagnostics Service Info 3 Pointer p-min.: -5.8 mbar p-max.: 167.5 mbar T-min.: -12.5 °C T-max.: +85.5 °C 3.
6 Adjustment with the indicating and adjustment module Service ▶ Basic adjustment Display Diagnostics Service Info Additional PA value 4 4.1 Secondary Value 1 Reset Out-Scale-Unit Volume l 4.3 Language Select reset ▼ Copy sensor data 4.2 4.4 Deutsch ▼ 4.7 Copy sensor data? PA-Out-Scale 100.00 lin % = 0.0 l 0.00 lin % = 100.0 l 4.3 SIL 4.5 Simulation 4.2 Start simulation ▼ HART mode Deactivated!▼ 4.6 Standard Address 0 4.
6 Adjustment with the indicating and adjustment module 6.7 Saving the parameter adjustment data It is recommended noting the adjusted data, e.g. in this operating instructions manual and archive them afterwards. They are hence available for multiple use or service purposes. If DPT10 is equipped with an indicating and adjustment module, the most important data can be read out of the sensor into indicating and adjustment module.
7 Setup with PDM 7 Setup with PDM 7.1 Parameter adjustment with PDM For WIKA sensors, instrument descriptions are also available as EDDs for the adjustment program PDM. The instrument descriptions are already implemented in the current version of PDM. For older versions of PDM, a free-of-charge download is available via Internet. 37245-EN-100120 Go via www.wika.com to the item "Service".
8 Set up 8 Set up 8.1 Select the mode The following modes can be adjusted on DPT10: l l l Flow measurement Level measurement Differential pressure measurement 8.2 Flow measurement In flow measurement, normally DPT10 is used without isolating diaphragm. Instructions Before adjusting DPT10, you have to clean the effective pressure lines and the instrument must be filled with the medium. Installation for gases 6 7 I + – 3 II 2 4 + A B Fig.
8 Set up Installation for liquids A + B 6 7 + – I III II 1 III 3 2 4 5 Fig. 45: Preferred installation for liquids I DPT10 II 3-fold valve block III Precipitator 1.5 Drain valves 2.4 Inlet valves 3 Breather valve 6.7 Vent valves on DPT10 A, B Blocking valves Prepare the adjustment Proceed as follows: 1 2 Close valve 3 Fill measuring system with medium.
8 Set up 4 Carry out a position correction if the following conditions apply. If the conditions are not fulfilled, then carry out the position correction after step 6. Conditions: The process cannot be sealed off. The pressure extraction points (A and B) are at the same geodesic height.
8 Set up Installation for open vessels Fig. 46: Preferred installation for open vessels I II 1 6.7 A Prepare the adjustment DPT10 Precipitator Drain valve Vent valves on DPT10 Blocking valve Proceed as follows: 1 Fill the vessel to just over the lower tap. 2 Fill measuring system with medium. 3 Vent instrument Open valve A: Medium flows in. Briefly open valve 6, then close it: Fill the measuring instrument completely with the medium and remove air.
8 Set up Installation for closed vessels B + A 6 7 + – I II III III 3 2 4 1 5 Fig.
8 Set up Installation for closed vessels with steam layer IV B + A 6 7 I + II III – III 3 2 4 1 5 Fig.
8 Set up Now: Valve 3, 6 and 7 are closed Valves 2, 4, A and B are open. Then carry out adjustment, see chapter "Set parameters". 8.4 Density and interface measurement For density and interface measurements, DPT10 with bilateral isolating diaphragm CSB is used. DPT10 in this version is immediately ready for operation. 8.5 Differential pressure measurement Instructions For differential pressure measurements, DPT10 without isolating diaphragm or with bilateral isolating diaphragm CSB is used.
8 Set up Installation for liquids A + B 6 7 + – I III II 1 III 3 2 4 5 Fig. 50: Preferred installation for liquids I DPT10 II 3-fold valve block III Precipitator 1.5 Drain valves 2.4 Inlet valves 3 Breather valve 6, 7 Vent valves on DPT10 A, B Blocking valves Prepare the adjustment Proceed as follows: 1 2 Close valve 3 Fill measuring system with medium. Open valves A, B, 2, 4: Medium flows in.
8 Set up Open valve 4: Connect minus side Now: Valves 1, 3, 5, 6 and 7 are closed6) Valves 2 and 4 open Valves A and B open (if available) Then carry out adjustment, see chapter "Set parameters". 68 Valves 1, 3, 5: Configuration with 5 valves.
9 Maintenance and fault rectification 9 Maintenance and fault rectification 9.1 Maintain Maintenance When the instrument is used properly, no special maintenance is required in normal operation. In some applications, product buildup on the separating diaphragms can influence the measuring result. Depending on the sensor and application, take precautions to ensure that heavy buildup, and especially a hardening thereof, is avoided. 9.
9 Maintenance and fault rectification ? Measured value on the indicating and adjustment module does not correspond to the value in the PLC l The menu item "Display - Display value" is not set to "PA-Out" à Check values and correct, if necessary ? No connection between PLC and PA network l Incorrect adjustment of the bus parameter and the segment coupler-dependent baud rate à Check data and correct, if necessary ? Instrument does not appear during connection setup l Profibus DP cable pole-reversed à C
9 Maintenance and fault rectification 9.3 Instrument repair You can download a return form (24 KB) in the Internet from our homepage www.wika.com under the item "Service".
10 Dismounting 10 Dismounting 10.1 Dismounting steps Warning: Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel, high temperatures, corrosive or toxic products etc. Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order. 10.2 Dispose The instrument consists of materials which can be recycled by specialised recycling companies.
11 Supplement 11 Supplement 11.1 Technical data General data Pressure type Differential pressure Measuring principle Piezoresistive Communication interface I²C bus Materials and weights Material 316L corresponds to stainless steel 1.4404 or 1.
11 Supplement Max. torque screws mounting strap 30 Nm Max. torque screws socket external housing 5 Nm (3.688 lbf ft) Weight approx. 4.2 … 4.5 kg (9.26 … 9.92 lbs), depending on process fitting Output variable Output signal digital output signal, format according to IEEE-754 Sensor address 126 (default setting) Current value 10 mA, ±0.5 mA Dynamic behaviour output ≤ 20 s Run-up time I 100 % 63 % t1 t2 t Fig.
11 Supplement Additional output parameter - temperature Processing is made via output signal HART multidrop, Profibus PA and Foundation Fieldbus Range -50 … +150 °C (-58 … +302 °F) Resolution 1 °C (1.8 °F) Accuracy in the range of 0 … +100 °C (+32 … +212 °F) ±3 K Accuracy in the range of -50 … 0 °C (-58 … +32 °F) and +100 … +150 °C (+212 … +302 °F) typ.
11 Supplement Influence of the installation position basic version ≤ 4 mbar9)10) A position-dependent zero-point shift can be corrected (see also chapter "General installation instructions" and "Installation instructions, isolating diaphragm systems"). Deviation determined according to the limit point method according to IEC 6077011) Applies to digital interfaces (HART, Profibus PA, Foundation Fieldbus) as well as to analogue current output 4 … 20 mA. Specifications refer to the set span.
11 Supplement Measuring range Thermal changes of the zero signal and the output span relating to the adjusted span -10 … +60 °C (+14 … +140 °F) 10 mbar, 30 mbar ±(0.31 x TD + 0.06) % 100 mbar ±(0.18 x TD + 0.02) % 500 mbar, 3 bar ±(0.08 x TD + 0.05) % 16 bar ±(0.1 x TD + 0.1) % 16 bar ±(0.08 x TD + 0.05) % 10 mbar, 30 mbar ±(0.45 x TD + 0.1) % 100 mbar ±(0.3 x TD + 0.15) % 500 mbar, 3 bar ±(0.12 x TD + 0.1) % 16 bar ±(0.15 x TD + 0.2) % 40 bar ±(0.37 x TD + 0.
11 Supplement Measuring cell 10 mbar 30 mbar 100 mbar 500 mbar Influence of the system pressure to the zero point ±0.28 % of URL/ 7 bar ±0.70 % of URL/ 70 bar ±0.42 % of URL/ 70 bar ±0.14 % of URL/ 70 bar Influence of the system pressure to the span ±0.28 % of URL/ 7 bar ±0.70 % of URL/ 70 bar ±0.42 % of URL/ 70 bar ±0.14 % of URL/ 70 bar Measuring cell 3 bar 16 bar 40 bar Influence of the system pressure to the zero point ±0.14 % of URL/7 bar ±0.14 % of URL/70 bar ±0.
11 Supplement Heating time - all versions ≤ 10 s Warm-up time Long-term stability (similar to DIN 16086, DINV 19259-1 and IEC 60770-1) Applies to digital interfaces (HART, Profibus PA, Foundation Fieldbus) as well as to analogue current output 4 … 20 mA. Specifications refer to the set span. Turn down (TD) = nominal measuring range/set span. Measuring cells ≥ 500 mbar ±0.05 % of the nominal measuring range/year, ±0.125 % of the nominal measuring range/5 years Measuring cells ≤ 500 mbar ±0.
11 Supplement Seal material Temperature limits FKM, cleaned for oxygen application -10 … +60 °C (-4 … +140 °F) PTFE, cleaned for oxygen application -20 … +60 °C (-4 … +140 °F) Process pressure limits Nominal range Nominal pressure Overload unilateral Overload bilateral 10 mbar (1 kPa) 160 bar (16000 kPa) 160 bar (16000 kPa) 240 bar (24000 kPa) 30 mbar (3 kPa) 160 bar (16000 kPa) 160 bar (16000 kPa) 240 bar (24000 kPa) 100 mbar (10 kPa) 160 bar (16000 kPa) 160 bar (16000 kPa) 240 bar (2
11 Supplement - Double chamber housing l 1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm), 1 x blind stopper M20 x 1.5; plug M12 x 1 for the external indicating and adjustment unit (optional) or: l 1 x closing cap ½ NPT, 1 x blind stopper ½ NPT, plug M12 x 1 for the external indicating and adjustment unit (optional) or: l Spring-loaded terminals for wire cross-section up to 1 x plug (depending on the version), 1 x blind stopper M20 x 1.
11 Supplement - Min. bending radius at 25 °C/77 °F 25 mm (0.985 in) - Diameter approx. 8 mm (0.315 in) - Colour Blue Cable entry/plug22) - External housing l 2 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x blind stopper M20 x 1.5 or: l Spring-loaded terminals for wire cross-section up to 1 x cable gland M20 x 1.5, 1 x plug (depending on version), 1 x blind stopper M20 x 1.5 2.
11 Supplement Electrical protective measures Protection rating - Housing, standard IP 66/IP 6723) IP 68 (1 bar)24) - Aluminium and stainless housing (optionally available) - Process component in IP 68 version IP 68 - External housing IP 65 Overvoltage category III Protection class II Approvals 37245-EN-100120 Depending on the version, instruments with approvals can have different technical data. For these instruments, please note the corresponding approval documents.
11 Supplement 11.2 Information on Profibus PA Instrument master file The instrument master file (GSD) contains the characteristic data of the Profibus PA instrument. These data are, e.g. the permissible transmission rates as well as information on diagnostics values and the format of the measured value outputted by the PA instrument. A bitmap file is also provided for the Profibus network planning tool. This file is installed automatically when the GSD file is integrated.
11 Supplement Sensor characteristics TB PROFIBUS PA-output Source for scaling Scaling Damping Alarms Failure mode ti Target Mode PA-Out °C FB 2 Fig.
11 Supplement Additional Cyclic Value l Byte-No. Format Value 1 2 3 4 IEEE-754Flieskommazahl PA-OUT (FB1) 5 Status Status (FB1) 6 7 8 9 IEEE-754Flieskommazahl Temperature (FB2) 10 Status Status (FB2) 11 12 13 14 IEEE-754Flieskommazahl Additional Cyclic Value 15 Status Status Fig. 54: Telegram configuration example 1 Example 2 with pressure value and temperature value without additional cyclical value: AI (PA-OUT) Temperature Free Place l l l Byte-No.
11 Supplement The status byte corresponds to profile 3.0 "Profibus PA Profile for Process Control Devices" coded. The status "Measured value OK" is coded as 80 (hex) (Bit7 = 1, Bit6 … 0 = 0). The measured value is transferred as a 32 bit floating point number in the IEEE-754 format.
11 Supplement Status code Description according to Pro- Possible cause fibus standard 0 x 89 good (non-cascade) - active ad- Lo-Alarm visory alarm - low limited 0 x 8a good (non-cascade) - active ad- Hi-Alarm visory alarm - high limited 0 x 8d good (non-cascade) - active crit- Lo-Lo-Alarm ical alarm - low limited 0 x 8e good (non-cascade) - active crit- Hi-Hi-Alarm ical alarm - high limited 37245-EN-100120 88 Differential pressure transmitter DPT10 - metal measuring diaphragm • Profibus PA
11 Supplement 11.3 Dimensions Plastic housing ~ 69 mm (2.72") ø 77 mm (3.03") ~ 87 mm (3.43") 112 mm (4.41") 120 mm (4.72") ø 84 mm (3.31") M20x1,5/ ½ NPT M20x1,5/ ½ NPT 1 1 2 2 Single chamber version Double chamber version Aluminium housing ~ 87 mm (3.43") ~ 116 mm (4.57") ø 84 mm (3.31") M20x1,5 M20x1,5/ ½ NPT 1 M20x1,5/ ½ NPT 2 Single chamber version Double chamber version 37245-EN-100120 1 2 120 mm (4.72") 116 mm (4.57") ø 84 mm (3.
11 Supplement Aluminium housing in protection rating IP 66/IP 68 (1 bar) ~ 105mm (4 9/64") ø 84mm (3 5/16") ~ 150mm (5 29/32") ø 84mm (3 5/16") M20x1,5 M20x1,5 M20x1,5/ ½ NPT 1 1 4 120mm (4 23/32") 116mm (4 9/16") M16x1,5 2 Single chamber version, Aluminium Double chamber housing Aluminium Version IP 68 with external electronics 103 mm (4.06") 68,3 mm (2.69") + – 59 mm (2.32") 93 mm (3.66") ø 55 mm (2.17") 1 90 mm (3.54") 110 mm (4.33") 115 mm (4.53") ~ 66 mm (2.6") ø 55 mm (2.
11 Supplement 41,3 mm (1.63") 85 mm (3.35") 134,5 mm (5.3") Oval flange, connection ¼-18 NPT or RC ¼ 7/16-20 UNF M 10 + – 1/4-18 NPT RC1/4 41,3 mm (1.63") 72 mm (2.84") 128 mm (5.04") 54 mm (2.13") 100 mm (3.94") 106 mm (4.17") 7/16-20 UNF M 10 (M12) + – 1/4-18 NPT RC1/4 100 mm (3.94") 54 mm (2.13") 98 mm (3.94") 37245-EN-100120 Fig. 63: Top: 10 mbar and 30 mbar measuring cell.
11 Supplement Version Connection Fastening Material Equipment U RC 1/4 7/16-20 UNF AISI 316L incl. 2 vent valves (316L) 1 1/4-18 NPT IEC 61518 PN 160: M10, PN 420: M12 Steel C 22.8 incl. 2 vent valves (316L) 2 1/4-18 NPT IEC 61518 PN 160: M10, PN 420: M12 AISI 316L incl. 2 vent valves (316L) 3 1/4-18 NPT IEC 61518 PN 160: M10, PN 420: M12 Alloy C276 without valves/closing screws + – 1/4-18 NPT RC1/4 54 mm (2.13") 100 mm (3.94") 106 mm (4.17") 42 mm (1.65") 41,3 mm (1.
11 Supplement Version Connection Fastening Material Equipment V RC 1/4 7/16-20 UNF AISI 316L incl. 4 closing screws (AISI 316L) and 2 ventilation valves 1 + – 54 mm (2.13") 98 mm (3.94") 41,3 mm (1.63") 128 mm (5.04") 72 mm (2.84") Oval flange, prepared for isolating diaphragm connection 7/16-20 UNF M 10 (M12) + – M6 2 3 54 mm (2.13") 98 mm (3.94") Fig. 65: left: Process fitting DPT10 prepared for isolating diaphragm connection.
Index INDEX A M Adjustment - Unit 46 - with differential pressure 48 - with ow 49-50 - with level 49 Application area - Density measurement 10 - Differential pressure measurement 9 - Flow measurement 8 - Interface measurement 10 - Level measurement 9 Maintenance 69 Mounting arrangement 27 Mounting position 14 O Oxygen applications 15 P Position correction 47 Process conditions 14 S C Connection compartment double chamber housing 36 D Density measurement 23 Differential pressure measurement - In gas
37245-EN-100120 Index Differential pressure transmitter DPT10 - metal measuring diaphragm • Profibus PA 95
Printing date:: WIKA Alexander Wiegand SE & Co. KG Alexander-Wiegand-Straße 30 63911 Klingenberg/Germany Phone +49 9372 132295 Fax +49 9372 132706 e-mail: support-tronic@wika.de www.wika.de All statements concerning scope of delivery, application, practical use and operating conditions of the sensors and processing systems correspond to the information available at the time of printing.