Owner's manual

- Turn down 1 : 1 up to 5 : 1 < 0.25 % x TD
- Turn down > 5 : 1 < 0.05 % x TD
Inuence of the product or ambient temperature
Thermal change zero signal and output span
Applies to the digital signal output (HART, Probus PA, Foundation Fieldbus) as well as to the
analogue current output 4 20 mA and refers to the set span. Turn down (TD) is the ratio nominal
measuring range/set span.
Thermal change zero signal and output span, reference temperature 20 °C (68 °F):
- In the compensated temperature range
0 +100 °C (+32 +212 °F)
< (0.05 % + 0.1 % x TD)
- Outside the compensated temperature
range
< (0.05 % + 0.15 % x TD)
Thermal change zero signal and output span with absolute pressure measuring range 0.1 bar,
reference temperature 20 °C (68 °F):
- In the compensated temperature range
0 +100 °C (+32 +212 °F)
< (0.1 % + 0.1 % x TD)
- Outside the compensated temperature
range
< (0.15 % + 0.15 % x TD)
Thermal change, current output
Applies also to the analogue 4 20 mA current output and refers to the set span.
Thermal change, current output < 0.15 % at -40 +80 °C (-40 +176 °F)
Long-term stability (similar to DIN 16086, DINV 19259-1 and IEC 60770-1)
Applies to digital interfaces (HART, Probus PA, Foundation Fieldbus) as well as to analogue
current output 4 20 mA. Specications refer to the set span. Turn down (TD) = nominal
measuring range/set span.
Long-term drift of the zero signal < (0.1 % x TD)/year
Ambient conditions
Ambient, storage and transport temperature -40 +80 °C (-40 +176 °F)
Process conditions
The speci cations to the pressure stage and the product temperature are used as an overview.
The speci cations of the type label are applicable.
Pressure stage, process tting
- Thread 316L PN 60
- Thread Aluminium PN 25
- Thread PVDF PN 10
- Hygienic ttings 316L PN 6, PN 10, PN 25, PN 40
- Flange 316L PN 16, PN 40 or 150 lbs, 300 lbs, 600 lbs
- Flange with extension 316L
Process pressure transmitter IPT-11 Vers. 4.0 - ceramic sensor 4 20 mA 41
10 Supplement
37301-EN-100302