User guide
1
2
Fig. 3: Position of the filter elements
1 Filter element for ventilation of the measuring cell
2 Filter element for ventilation of the electronics housing
Information:
Make sure that the filter elements are always free of buildup during
operation. A pressure washer must not be used for cleaning.
Higher process temperatures mean often also higher ambient
temperatures.
1
2
Fig. 4: Temperature ranges
1 Process temperature
2 Ambient temperature
Make sure that the upper temperature limits for the environment of
electronics housing and connection cable specified in chapter
"Technical data" are not exceeded.
4.2 Mounting instructions
IPT-1* Vers. 2.0 functions in any installation position. It is mounted
according to the same directives as a manometer (DIN EN 839-2).
2)
2)
Probable position correction see chapter "Setup steps".
Temperature limits
Mounting position
Process pressure transmitter IPT-10 Vers. 2.0 - ceramic sensor • Profibus PA 11
4 Mounting
31541-EN-081211