Manual

Indicating and adjustment module
Voltage supply and data transmission through the sensor
Indication LC display in dot matrix
Adjustment elements 4 keys
Protection
- unassembled IP 20
- mounted into the sensor without cover IP 40
Materials
- Housing ABS
- Inspection window Polyester foil
Power supply
Operating voltage
- Non-Ex instrument 12 36 V DC
- EEx-ia instrument 12 30 V DC
Operating voltage with lighted indicating and adjustment module
- Non-Ex instrument 20 36 V DC
- EEx-ia instrument 20 30 V DC
Permissible residual ripple
- < 100 Hz U
ss
< 1 V
- 100 Hz 10 kHz U
ss
< 10 mV
Load see diagram
1000
750
500
250
12 18
1614 20 22 24 26 28 30 32 34 36
V
4
1
2
3
Fig. 20: Voltage diagram
1 HART load with HART version
2 Voltage limit EEx-ia instrument
3 Voltage limit non-Ex instrument
4 Operating voltage
Process pressure transmitter IPT-10 Vers. 2.0 - ceramic sensor 4 20 mA/HART 49
10 Supplement
31540-EN-090130