Owner's manual

Nominal range Overload capacity, max.
pressure
Overload capacity, min.
pressure
-15 25 psig 700 psig -15 psig
-15 70 psig 950 psig -15 psig
-15 150 psig 1300 psig -15 psig
-15 350 psig 1900 psig -15 psig
-15 900 psig 2900 psig -15 psig
-0.7 0.7 psig 200 psig -3 psig
-1.5 1.5 psig 290 psig -6 psig
-3 3 psig 430 psig -12 psig
-7 7 psig 500 psig -15 psig
Absolute pressure
0 1.5 psig 200 psig 0 psig
0 15 psig 500 psig
0 psig
0 35 psig 700 psig
0 psig
0 70 psig 900 psig
0 psig
0 150 psig 1300 psig
0 psig
0 350 psig 1900 psig
0 psig
0 900 psig 2900 psig
0 psig
Reference conditions and actuating variables (similar to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
- Temperature +15 +25 °C (+59 +77 °F)
- Relative humidity 45 75 %
- Air pressure 860 1060 mbar/86 106 kPa (12.5 15.4 psig)
Determination of characteristics Limit point adjustment according to IEC 61298-2
Characteristics linear
Reference installation position upright, diaphragm points downward
Inuence of the installation position < 0.2 mbar/20 Pa (0.003 psig)
Deviation determined according to the limit point method according to IEC 60770
9)
Applies to digital interfaces (HART, Probus PA, Foundation Fieldbus) as well as to analogue
current output 4 20 mA. Specications refer to the set span. Turn down (TD) is the ratio nominal
measuring range/set span.
Deviation
- Turn down 1 : 1 up to 5 : 1 < 0.075 %
- Turn down > 5 : 1 < 0.015 % x TD
Deviation with absolute pressure measuring range 0.1 bar
9)
Incl. non-linearity, hysteresis and non-repeatability.
Process pressure transmitter IPT-10 Vers. 2.0 - ceramic sensor 4 20 mA 39
9 Supplement
37299-EN-100302