User guide

Deviation with absolute pressure measuring range 0.1 bar
- Turn down 1 : 1 up to 5 : 1 < 0.25 %
- Turn down > 5 : 1 < 0.05 % x TD
Inuence of the product or ambient temperature
Applies to instruments with digital signal output (HART, Probus PA, Foundation Fieldbus) as well
as to instruments with analogue current output 4 20 mA. Specications refer to the set span.
Turn down (TD) = nominal measuring range/set span.
Thermal change zero zignal, reference temperature 20 °C (68 °F):
- In the compensated temperature range
0 +100 °C (+32 +212 °F)
< 0.05 %/10 K x TD
- Outside the compensated temperature
range
typ. < 0.05 %/10 K x TD
Applies also to instruments with analogue 4 20 mA current output and refers to the set span.
Thermal change, current output < 0.15 % at -40 +80 °C (-40 +176 °F)
Long-term stability (similar to DIN 16086, DINV 19259-1 and IEC 60770-1)
Applies to digital interfaces (HART, Probus PA, Foundation Fieldbus) as well as to analogue
current output 4 20 mA. Specications refer to the set span. Turn down (TD) = nominal
measuring range/set span.
Long-term drift of the zero signal < (0.1 % x TD)/year
Ambient conditions
Ambient, storage and transport temperature
- Standard version -40 +80 °C (-40 +176 °F)
- Standard version with indicating and
adjustment module
-15 +70 °C (+5 +158 °F)
- Connection G1 A front-ush according to
EHEDG
-10 +80 °C (+14 +176 °F)
- Fitting G1 A front-ush according to
EHEDG with indicating and adjustment
module
-10 +70 °C (+14 +158 °F)
- Version for oxygen applications
15)
-40 +60 °C (-40 +140 °F)
- Version for oxygen applications with
indicating and adjustment module
16)
-15 +60 °C (+5 +140 °F)
Process conditions
The specications are used as an overview. For pressure stage and product temperature the
specications on the type plate are applicable.
15)
Up to 60 °C (140 °F).
16)
Up to 60 °C (140 °F).
42 Process pressure transmitter IPT-1* Vers. 3.0 - metalic sensor Foundation Fieldbus
10 Supplement
31545-EN-081211