User Manual

45
7Probe
7.1 Probe description
7.1.1 Probe type
L-Linear array probe
C-Convex array probe
E-Cavity probe.
7.1.2 Probe composition
The probe is composed of piezoelectric ceramic wafer, acoustic focusing lens, shell and so on.
7.1.3 Probe performance index
The performance indicators of each probe are published as follows:
Table 1 Image performance Indexes
Mode
Order
number
Project
Probe model
C5-2Sv C5-2v L11-4v L11-4Sv
B/M
A
Probe nominal
frequency (MHz)
3.5 3.5 7.5 7.5
B
Deviation
between acoustic
working
frequency and
nominal
frequency(%)
≤±
15
≤±
15
≤±
15
≤±
15
C
Maximum
detection
depth(mm)
180
170
60
60
D
lateral resolution
(mm)
2(depth
80)
3(80
depth
130)
2(depth
80)
3(80
depth
130)
2
(depth
40)
2
(depth
40)
E
Azimuthal
resolution (mm)
1(depth
80)
2(80
depth
130)
1(depth
80)
2(80
depth
130)
1
(depth
50)
1
(depth
50)
F Fade zone(mm)
3
3
3
3
G
Slice
thickness(mm)
8
8
8
8