User Manual

46
Mode
Order
number
Project
Probe model
C5-2Sv C5-2v L11-4v L11-4Sv
H
Transverse
geometric position
precision(%)
10
10
5
10
I
Longitudinal
geometric position
precision(%)
10
10
5
5
J
Perimeter and
area measurement
deviation(%)
≤±
15
≤±
15
≤±
15
≤±
15
K
M mode time
display
deviation(%)
10
10
10
10
Color
L
Service frequency
MHz
2.0 3.5 2.5 3.5 6.0 7.5 6.0 7.5
M
Depth of
investigation
(mm)
120 120 40 40
N
Relationship
between color
blood flow image
and gray-scale
image of the pipe
in which it is
located
coincide coincide coincide coincide
O
Blood flow
direction
requirement
It can be correctly
identified and there is
no aliasing.
It can be correctly
identified and there is
no aliasing.
It can be
correctly
identified and
there is no
aliasing.
It can be
correctly
identified and
there is no
aliasing.
PW
P
Service frequency
MHz
2.0 3.5 2.5 3.5 6.0 7.5 6.0 7.5
Q
Depth of
investigation(mm)
120
120
40
40
R
Veloci ty
measurement
deviation(%)
±
20
±
20
±
20
±
20