User Manual

48
Mode
Order
number
Project
Probe model
E10-4v C8-5v
Color
L
Service frequency
MHz
5.0 6.0 5.0 6.0
M
Depth of
investigation (mm)
40
40
N
Relationship
between color blood
flow image and
gray-scale image of
the pipe in which it
is located
coincide coincide
O
Blood flow
direction
requirement
It can be correctly identified
and there is no aliasing.
It can be correctly identified and there
is no aliasing.
PW
P
Service frequency
MHz
5.0 6.0 5.0 6.0
Q
Depth of
investigation(mm)
40
40
R
Velocity
measurement
deviation(%)
±20 ±20
S
Accuracy of
position of Vernier
in Doppler
sampling area
Near the wall of the pipe, no
velocity indication; center of
pipe, maximum velocity
display
Near the wall of the pipe, no velocity
indication; center of pipe, maximum
velocity display
7.2 Probe description
In order to reduce the spread of the disease, it is necessary to take some protective measures.
In clinic, the probe sheath is conducive to prevent infection. Strongly recommended in accordance
with requirements of the sterile transducer sheath cavity inspection.
On the animal body are examined using the probe to probe must be put on the probe set.
Please use the protection cover market acceptance.
CAR
EFUL
1. In order to avoid infection, in the inspection, the probe cover
can only be used once.
2. The probe sheath is made of natural latex and talc, which may cause
allergic reactions in some people
3. Do not use expired probe sheath. Check the probe sheath for
expiration before using the probe sheath
Operation procedure (for reference only)