User Manual
49
1.In the protective sleeve or inside the probe surface coated with an appropriate amount of
coupling agent, without the use of coupling agent, the image is not clear.
2.To probe into the protective sleeve to ensure the sterility of the probe. The protective sleeve
of the tension to remove wrinkles and bubbles, do not pull too far.
3.With the rope to ensure the protection of security.
4.Check the protective sleeve to ensure no damage
7.3 Inspection and maintenance
7.3.1 Inspect
Frequently check the probe cable, if found damaged, broken phenomenon, prohibit the use of
immediate replacement or repair.
Regularly check the socket, the sound window parts, if found damaged, bubble phenomenon,
prohibit the use of immediate replacement or repair.
7.3.2 Life cycle
According to the manufacturer's design, production and other related documents, this type of
product life is generally 5 years, constituting the product material over time will gradually aging,
continue to use the products beyond the life of the post, may cause performance degradation and
failure rate is significantly high.
7.3.3 Probe maintenance
The probe after every use, with a soft cloth with a damp heat 3.4% saturated aldehyde
solution gently and thoroughly probe.
The probe is expensive wearing parts, no collision, drop. When diagnosis is suspended, it
should be placed in the probe box, and the instrument under the "frozen" state.
Medical ultrasonic coupling agent should be used in diagnosis.
CAREFU
L
:
The probe on the probe for a long time with a coupling agent may
damage the probe.
WAR
NING
:
The manufacturer will not be held responsible for the risks arising
from the continued use of the product life cycle.