The probe must be cleaned after each use;
2
.
Do not use surgical brush to clean the probe, even if the use of soft
brush may damage the probe, only use a soft cloth;
WARNIN
G
:
1
.
The product must be away from machine with high current or
strong magnetic.
2
.
Do not immerse the probe in any type of liquid or detergent.
3
.
Probe is strictly prohibited from immersing by any type of
liquid.
4
.
The method of disinfecting probe with gas or heating is
strictly prohibited.
5
.
It is strictly prohibited to probe cable probe socket or
immersed in the solution, although the surface of the probe
is waterproof but only the head of the probe. Please be careful
using the coupling agent and cleaning, disinfection of the
probe surface.
6
.
Equipment failure, Users are not allowed to repair without
authorization. Product must be sent back to the company
Check the following items after cleaning and disinfection:
Inspection
item item item
Inspection method Acceptance Criteria
Lens Visual inspection of the lens surface and the
connection between the lens and the shell of the
acoustic head
There shall be no cracks in the
connection between the lens and the
housing.
Shell Visual inspection of shell surface No damage or crack on the surface o
f
the shell
Shell
connection
Visual inspection of shell connection There shall be no cracks in the joint
of the shell, lack of glue
Hangover Visual inspection of probe surfaces, especiall
y
pits and narrow areas
There shall be no residual detergent
or disinfectant