User Manual

52
8 Puncture guide
8.1 Enter or exit puncture mode
Under the Windows platform, in the SoftKey area, click B mode [Image Parameter Settings],
and select [puncture (on / off)].
Set [Piercing frame model] for quick calibration.
Set [guide line], depending on needle selection, A, B, and ALL.
Set [linetype], have thick, medium, thin.
Exit puncture guiding function
The system is in the puncture status. Click[Exit] to exit the puncture status.
8.2 Puncture lead calibration
Place the puncture frame and the sound head of the puncture needle in the water;
Click [calibrate] to align the needle with the puncture guide line by adjusting the “position
“angle” “+” and“-”.
After coincident, save the calibrated puncture guide wire parameters.
WAR
NING
1. The guide leads must be calibrated before each puncture.
2. The frozen is not allowed at the time of puncture.
3. If the puncture needle is not consistent with the position of the
puncture guide lead, do not puncture.