If the rated maximum output power of the transmitter is not included in the values given above, can
be estimated by using the corresponding column isolation distance equation. In the equation P is the
nominal maximum output power of transmitter manufacturers are given, measuredinwatts.
If the system image severely disturbed affect diagnosis, it is necessary to set the device away from
the noise source or install the external RF transmission power noise filter the noise reduced to an
acceptable level.
Be careful 1 : In 80MHz-800MHz, using a higher frequency formula
Be careful 2: These guidelines do not apply in all cases. The material structure, objects and
people can absorb and reflect electromagnetic waves, thus affecting the electromagnetic
propagation.
Table 5
Fundamental characteristics
Scanning mode: the mode of operation of an ultrasonic diagnostic device, including a series
of ultrasonic pulses that generate scanning lines that do not follow the same acoustic path.
Scanning mode Parameter
Bmode
Gain:0-100 tunable
Maximal acoustic output :10-100% tunable
Depth:
—
C5-2Sv probe
≥
30cm
—
C5-2v probe
≥
30cm
— L11-4v probe≥14cm
—
L11-4Sv probe
≥
14cm
—
C8-5v probe
≥
14cm
— E10-4v probe≥14cm
Mmode
Cmode
Power mode
PW mode