Instruction Manual

Wireless TENS/EMS RS-18 Instruction Manual
26
Information about Electromagnetic Compatibility
The RS-18TENS/EMS Remote Control and Pod(s) are intended for use in typical surroundings that have
been authorized according to the EN/IEC 60601-1-2:2014+A1:2020 EMC safety standard.
This device meets the CISPR standard, which means that radio frequency (RF) emissions are unlikely to
cause interference with surrounding electronic equipment (radios, computers, telephones, etc.).
The RS-18TENS/EMS Remote Control and Pod(s) are built to withstand electrostatic discharge, magnetic
fields from the mains power source, and radio frequency transmitters.
However, it is impossible to guarantee that the stimulator would not be damaged by powerful RF (radio
frequency) forces emanating from other sources.
Refer to the EMC tables for more complete information on electromagnetic emissions and immunity.
EMMISSION:
STANDARD ITEM REMARKS
CISPR 11: 2011 Conducted
Radiated
IEC 61000-3-2:2014 Harmonic current emissions
IEC 61000-3-3:2013
IMMUNITY:
STANDARD ITEM IEC 60601-1-2 Test
Levels for Home
Healthcare Environment
RS-18 TENS/EMS
Stimulator Test
Levels
REMARKS
IEC 61000-4-
2:2008
ESD 3 8 kV contact;
3 2 kV, 3 4 kV, 3 8 kV, 3
15 kV air
IEC61000-4-
3:2010
RS 10 V/m
80 MHz 2.7 GHz
80% AM at 1 kHz
IEC 61000-4-
4:2012
EFT
3 2 kV
100 kHz repetition
frequency
IEC 61000-4-
5:2014
Surge 3 0.5 kV, 3 1 kV
IEC 61000-4-
6:2013
CS 3V 0.15 MHz 80 MHz
6V in ISM and amateur
bands between 0.15
MHz and 80 MHz