Wireless TENS/EMS RS-18 Instruction Manual 
26 
Information about Electromagnetic Compatibility 
The RS-18TENS/EMS Remote Control and Pod(s) are intended for use in typical surroundings that have 
been authorized according to the EN/IEC 60601-1-2:2014+A1:2020 EMC safety standard. 
This device meets the CISPR standard, which means that radio frequency (RF) emissions are unlikely to 
cause interference with surrounding electronic equipment (radios, computers, telephones, etc.). 
The RS-18TENS/EMS Remote Control and Pod(s) are built to withstand electrostatic discharge, magnetic 
fields from the mains power source, and radio frequency transmitters. 
However, it is impossible to guarantee that the stimulator would not be damaged by powerful RF (radio 
frequency) forces emanating from other sources. 
Refer to the EMC tables for more complete information on electromagnetic emissions and immunity. 
EMMISSION: 
STANDARD  ITEM  REMARKS 
CISPR 11: 2011  Conducted   
Radiated   
IEC 61000-3-2:2014  Harmonic current emissions   
IEC 61000-3-3:2013     
IMMUNITY: 
STANDARD  ITEM  IEC 60601-1-2 Test 
Levels for Home 
Healthcare Environment 
RS-18 TENS/EMS 
Stimulator Test 
Levels 
REMARKS 
IEC 61000-4-
2:2008 
ESD  3 8 kV contact; 
3 2 kV, 3 4 kV, 3 8 kV, 3 
15 kV air 
IEC61000-4-
3:2010 
RS  10 V/m 
80 MHz – 2.7 GHz 
80% AM at 1 kHz 
IEC 61000-4-
4:2012 
EFT 
3 2 kV 
100 kHz repetition 
frequency 
IEC 61000-4-
5:2014 
Surge  3 0.5 kV, 3 1 kV     
IEC 61000-4-
6:2013 
CS  3V 0.15 MHz – 80 MHz 
6V in ISM and amateur 
bands between 0.15 
MHz and 80 MHz   