There are two types of limits, Absolute and Relative.  The LDM bit in the CMM register 
establishes  which  type  will  be  used,  where  “0”  indicates  Absolute  and  “1”  indicates 
Relative. In Absolute  Alarm  Mode,  the  limits  are  fixed  and  do  not  change,  while  in 
Relative Alarm Mode the limits change whenever the current Alarm Limits are exceeded. 
As  the  name  suggests,  Absolute  Alarm  Mode  is  used  for  monitoring  the  absolute 
magnetic field, while Relative Alarm Mode is used to monitor changes in magnetic field. 
Absolute Alarm Mode 
In Absolute Alarm Mode the limits are set for each axis by the Alarm Lower Limit 
and  Alarm  Upper  Limit  registers,  as  given  in  Table  5-5.    The  register  values  are 
unsigned integers.  The ALARM bit will go HIGH if the measured value (MX, MY, 
or MZ) on any given axis is either greater than the Alarm Upper Limit for that axis or 
less than the Alarm Lower Limit for that axis.  This is an “OR” function, looking at 
both the Alarm Upper and Lower Limits, and looking at all sensors being measured. 
Table 5-5: Alarm Lower and Upper Limit Registers 
Register Description 
Register 
Name 
Read 
(Hex) 
Write 
(Hex) 
X Axis Alarm Lower Limit - MSB 
ALLX2 
8C 
0C 
X Axis Alarm Lower Limit - mid 
ALLX1 
8D 
0D 
X Axis Alarm Lower Limit - LSB 
ALLX0 
8E 
0E 
X Axis Alarm Upper Limit - MSB 
AULX2 
8F 
0F 
X Axis Alarm Upper Limit - mid 
AULX1 
90 
10 
X Axis Alarm Upper Limit - LSB 
AULX0 
91 
11 
Y Axis Alarm Lower Limit - MSB 
ALLY2 
92 
12 
Y Axis Alarm Lower Limit - mid 
ALLY1 
93 
13 
Y Axis Alarm Lower Limit - LSB 
ALLY0 
94 
14 
Y Axis Alarm Upper Limit - MSB 
AULY2 
95 
15 
Y Axis Alarm Upper Limit - mid 
AULY1 
96 
16 
Y Axis Alarm Upper Limit - LSB 
AULY0 
97 
17 
Z Axis Alarm Lower Limit - MSB 
ALLZ2 
98 
18 
Z Axis Alarm Lower Limit - mid 
ALLZ1 
99 
19 
Z Axis Alarm Lower Limit - LSB 
ALLZ0 
9A 
1A 
Z Axis Alarm Upper Limit - MSB 
AULZ2 
9B 
1B 
Z Axis Alarm Upper Limit - mid 
AULZ1 
9C 
1C 
Z Axis Alarm Upper Limit - LSB 
AULZ0 
9D 
1D 