User Manual
PNI Sensor Corporation  Doc 1017252 r04 
RM3100 & RM2100 Sensor Suite User Manual  Page 36 of 45 
Table 5-8: Measurement Results Registers 
Register Description 
Write Address 
(Hex) 
Read Address 
(Hex) 
X Axis Measurement (2) 
24 
A4 
X Axis Measurement (1) 
25 
A5 
X Axis Measurement (0) 
26 
A6 
Y Axis Measurement (2) 
27 
A7 
Y Axis Measurement (1) 
28 
A8 
Y Axis Measurement (0) 
29 
A9 
Z Axis Measurement (2) 
2A 
AA 
Z Axis Measurement (1) 
2B 
AB 
Z Axis Measurement (0) 
2C 
AC 
Normally  it  is  only  necessary  to  send  “A4
H
”,  since  the  register  value  automatically 
increments  on  the  clock  cycles  such  that  after  sending  “A4
H
”  all  3  bytes  for  the  X  axis 
measurement would be clocked out, then the 3 bytes for the Y axis measurement, then the 3 
bytes for the Z axis measurement. After these 9 bytes have been clocked out, the subsequent 
output data has no relevance. 
5.6  Troubleshooting and General Information 
The MagI2C incorporates several registers that can be used to help troubleshoot the system 
or device.  These include the Built-In Self Test (BIST) register, the Handshake register, and 
the RevID register. 
5.6.1  Built-In Self Test Register (0x33) 
The  Built-In  Self  Test  (BIST)  register  checks  the  status  of  the  internal  LR  circuit 
oscillator.  Write to the BIST register by sending 0x33  followed by the BIST register 
values, and request a read from the BIST register by sending 0xB3. The register contents 
are defined below. 
Bit # 
7 
6 
5 
4 
3 
2 
1 
0 
Value 
STE 
ZOK 
YOK 
XOK 
BW1 
BW0 
BP1 
BP0 
` 