Instruction Manual
EMC Tables
21
Recommended separation distances between portable and mobile RF
communications equipment and the GSH822 Blood Pressure Monitor
The GSH822 Blood Pressure Monitor is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the
user of the GSH822 Blood Pressure Monitor can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
communication equipment (transmitters) and the GSH822 Blood Pressure Monitor
as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power
of transmitter
W
150 kHz to 80 MHz
d= 1.2 P
80 MHz to 800 MHz
d= 1.2 P
800 MHz to 2.5 GHz
d= 2.3 P
Separation distance according to frequency of transmitter
m
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guildelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
0.12
0.38
1.2
3.8
12
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
0.01
0.1
1
10
100