Instruction manual

12 World Precision Instruments
Gasket Color Green Black Red White
Pipette Diameter (mm) 1.0 1.2 1.5 1.65
3. Connect the other end of the small tubing to the appropriate port on the front of the
PicoPump.
!
WARNING: DANGER OF INJURY EXISTS IF THE PIPETTE IS INSECURE.
HIGH PRESSURE CAN CAUSE EJECTION AT HIGH VELOCITY.
NOTE: Long lengths of tubing tend to decrease the time response of the system due to
a loading effect of the increased volume. For fastest possible response, keep the tubing
length (and diameter) as small as possible.
OPERATING INSTRUCTIONS
!
CAUTION: Be careful NOT apply too much pressure when turning the knobs fully
counterclockwise to minimum pressure. They cannot be adjusted to 0.0 PSI.
Over-tightening the knobs damages the regulators.
!
CAUTION: The regulators leak a small amount of gas as a part of their normal
operation. To conserve tank contents, make sure to turn off the main gas supply
pressure when the PicoPump is not in use.
Techniques In Microinjection
The PicoPump was designed for demanding tasks like the microinjection of fl uid into
cells. In this section, we will look at several important things to keep in mind when
working with the PicoPump.
Setting the Hold Pressure
The Hold pressure is used to counterbalance the capillary action of the fl uid backfi lling
into the pipettes. If you insert an empty pipette into fl uid, you can see a meniscus rising
from the capillary tip.
1. Set the Solenoid Input Selector Switch to the Hold position.
2. Set the Hold pressure by adjusting the regulator setting until the meniscus stops
at the desired position. In many applications, colored dye or fl uorescent dye is
dissolved in the injection fl uid. The capillary effect may be observed with the color
change at the tip of the pipette. When the fl uid fl ows into the pipette, the color of
the tip becomes lighter. If the hold pressure is higher than the capillary pressure,
the fl uid oozes out of the pipette. The solution around the pipette will be colored.
Adjusting the Hold pressure prevents this from happening.
Understanding how capillary action causes the front fi lling of the pipette helps you to
correctly use the Hold pressure. The fl ow rate is determined by the capillary action and
the tip size. Since the tip size is often determined by the requirement of the application,
controlling the hold pressure becomes the main option to eliminate the uptake of fl uid by
capillary action.