Instruction manual

PV820/PV830 Pneumatic PicoPump
World Precision Instruments 15
For ejected volumes less than 1nL, visual inspection in air proves to be diffi cult due
to rapid evaporation. The same technique may be used though if the drop is kept
submerged under oil. Droplets may seem to disappear after emergence from the tip.
Sometimes this is due to creepage of the aqueous fl uid back along the outside shank of
the micropipette. This creepage may be decreased by silanizing the outside of the pipette.
If you choose to silanize only the outside of the pipette, the PicoPump may be used to
applying air fl ow through the pipette during the silanization.
TIP: Precise assays of ejected volume may also be obtained through various
radioisotopic methods. See the bibliography for further information.
A slight defl ection of the micropipette tip may be noticed during the application of the
pressure pulse. This defl ection may be eliminated by ensuring that the micropipette is
rmly seated in the holder and that the holder is fi rmly held by a manipulator.
Rate Calibration using a Known Volume
We can calculate the volume of liquid contained in a 1mm length of a glass pipette, if
we know the inner diameter (ID). Then, we can inject that known volume and measure
the time it takes to inject it. From this we can calculate the rate of injection. For example,
if the volume in 1mm of glass with 0.58μm ID (WPI #1B100) is 264nL, and it takes 30
seconds to inject that volume, then the fl ow for 1 second is 264nL/30s or 8.8nL/second,
and a 1nL volume then takes 0.113 seconds to inject.
This method is discussed in the following JoVE video: http://www.jove.com/video/2079/
intravenous-microinjections-zebrafi sh-larvae-to-study-acute-kidney.
In the Jove video, the capillary tip is 10–20μm. By following these steps, we are able to
calculate the timing needed to deliver a 1nl injection by counting the time between 1mm
marks as the fl uid is injected.
NOTE: The PicoPump timing should not be adjusted to go below 10–15ms as an
absolute minimum, since it takes 6–10ms for the pressure relay to respond.
1. Fill a small dish with mineral oil and place it under a stereo microscope.
2. Turn on your air and vacuum pumps, and set your hold pressure and your eject
pressure. The proper settings must be determined experimentally. Turn on the
PicoPump.
NOTE: The eject pressure must exceed the hold pressure in order to eject the fl uid.
3. Use a pipettor to inject a 10μl sphere of fl uid into the mineral oil.
4. Using a fi ne pointed permanent marker, mark the injecting micropipette at 1mm
increments.
5. Mount your “graduated” micropipette on the PicoNozzle that is connected to the
Eject port on the PicoPump.
6. Submerge the tip of the pipette in the fl uid droplet you placed in the mineral oil.
Front fi ll the injecting pipette using the Vac (PV830)/Vent (PV820) setting on the