Instruction manual
USER INTERFACE
2-5
Base Entry
The Base value is usually set to 0% for +DOT and 100% for -DOT, or
adjusted to a value equal to a known minimum (base) dot size (ex. 3%
for +Dot or 97% for −Dot).
1. Repeatedly depress [
FUNCTION] key until "+Dot" or "−Dot" is
displayed.
2. Select color by depressing [
COLOR].
3. Hold [
ZERO] key down (thru step 4).
• Base value is displayed.
4. Enter Base value using the [
FUNCTION] key to decrease or the
[
COLOR] key to increase value. (depress both to reset to zero.)
Release keys.
5. Depress [
FUNCTION] to return to normal operation.
Measurement Procedure - General
1. Select desired function and color.
2. Center area being measured over center of aperture.
3. Lower reading head by pressing on Read button. "READING" will
be displayed during a measurement cycle.
4. Release Read Button after data is displayed.
NOTE: "INVALID" will be displayed if the Read button is not
depressed for enough time.
Display Messages
MESSAGE REASON
BATTERY: The memory backup battery has failed. The Lithium
battery needs replacement by a qualified technician.
INVALID: The Read button was held down too short causing an
invalid measurement. If “INVALID” appears after the
Read button was held down for the proper length of
time or if “READING” is displayed an abnormal
length of time while taking a measurement, possible
causes are defective Side Sensor, Reading Head
Assembly, or Transmission PCB.
LAMP FAIL: The Read Lamp has failed its intensity test. The Read
Lamp should be examined and possibly replaced.
When this happens, you can get out of this error.