MESSAGE REASON
condition (if lamp was OK or replaced) by pressing
[FUNCTION], then [COLOR], then [FUNCTION].
MEM TEST: The memory in the unit is going through an extended
memory check. If “MEM TEST” remains on the
display, the display PCB needs to be replaced.
MEM LOST: Calibration of the unit has been lost and recalibration
is necessary.
NEED CAL: Unit needs full length calibration.
uP FAIL: The microprocessor has failed its memory test. The
display PCB should be replaced by a qualified
technician.
311 Emulation
When 311 EMULATION is set to ON, the 361T will simulate the I/O
port and some operational characteristics of the X-Rite 311/RS
densitometer. The modes are setup as follows: x10 OFF, I/O Port [RCI
ON, RPT OFF, P5 OFF, BAUD 1200, HDR OFF, DPT OFF, CR,
COMP OFF, and X OFF], SEQUENCE OFF, and READ LAMP ON.
The operational characteristics are setup as follows: DENSITY
operation is the only function that is accessible, and MODES are
disabled.
NOTE: To access 311 emulation, the AC power must be
turned off and then turned back on with the function and
color keys held depressed thru the self test until "311 off"
is displayed.
Described below is the procedure for setting the 311 Emulation.
1. Turn OFF power to the 361T.
2. Depress [
FUNCTION] and [COLOR] keys together (than turn on
A.C Power) until "311 OFF" is displayed.
3. Each depression of the [
ZERO] key will alternate between 311 ON
and 311 OFF.
4. Depress [
FUNCTION] one time to return to normal operation.