Instruction manual

CHAPTER FIVE
5-4
Setting the Sequence Structure
Sequence allows you to setup a measurement routine that fits your
individual quality control requirements. Sequence has a maximum of 36
steps and a function for each step can be selected.
SQ OFF/ON is used to disable or enable SEQUENCE from
appearing as an active function in the main menu.
COL ADJ/LOC is used to allow ADJustment of what color
response is used during sequence or to LOCk sequence into one
response only.
SQ order/class/each is used to determine the method for printout of
data.
1. order – Transmit all data by sequential step number.
2. class – Transmit data by type of function. (i.e. DEN first,
then +Dot, then Dot)
3. each – Transmit data after each step is measured.
SIZE 1-36 sets the size of the sequence from 1 to 36 steps.
FUNCTION Den/+Dot/Dot/Skip sets the function that will be
used for each sequence step.
NOTE: SKIP is used for steps that you don’t want to measure,
but you still want the step number (during measurement and
print out) to correlate to the actual step numbers of the film. The
steps you set as SKIP will not be displayed during Sequence
operation; instead the display will skip over them and designate
the actual step you are support to read. For example, if you set
steps 2 and 5 as SKIP, your 361T will display step 1, 3, 4, 6, 7,
etc. as you advance thru sequence.