Instruction manual
MODE SELECTION
5-5
The procedure for setting Sequence structure is shown below.
1. Depress [
FUNCTION] and [COLOR] keys at the same time until "N
DcalY" is displayed.
2. Depress [
FUNCTION] for NO.
3. Depress [
FUNCTION] for NO.
4. Depress [
ZERO] for Yes.
5. Depress [
FUNCTION] two (2) times "↓ SEQ Y" is displayed.
6. Depress [
ZERO] to advance to SQ. Each depression of [ZERO] will
alternate between SQOFF and SQ On.
NOTE: If "←SQ OFF" is displayed, depressing the
FUNCTION key will cause the display to return to Step 4.
7. Depress [FUNCTION] to advance to COL. Each depression of
[
ZERO] will alternate between COL LOC and COL ADJ.
8. Depress [
FUNCTION] to advance to SQ. Each depression of [ZERO]
will page thru SQ class, SQ each, SQ order.
9. Depress [
FUNCTION] to advance to SIZE. Hold dc (ZERO)
pressed down and press d (FUNCTION) to decrease or c (COLOR)
to increase the amount of steps required (1 thru 36).
10. Depress [
FUNCTION] to advance to DEN 00 (Step 00).
11. Repeatedly press the [
COLOR] key until "←SKIP 00" is displayed.
(This will cause Step 00 to be skipped during operation.) Or select
[←ZERO 00] to enable a special step 0 for zeroing the
densitometer to the film base.
12. Hold dc (
ZERO) pressed down and press c (COLOR) once to
advance to Step 1. Release keys.
13. Repeatedly press [
COLOR] key to select the function (DEN, +DOT,
-DOT, or SKIP) that will be required for the step.
14. Hold dc (
ZERO) pressed down and press c (COLOR) once to
advance to Steps 2, 3, 4, etc.
15. Depress [
FUNCTION] three (3) times and 361T returns to normal
operation.