Instruction manual

CHAPTER SIX
6-6
T [BAUD] is used to change the baud rate of the 361T. The unit must
receive the following as the address and the action codes portion of
the command string:
XXT
—— Action Code
———— 03 = 300 baud
06 = 600 baud
0C = 1200 baud
18 = 2400 baud
30 = 4800 baud
60 = 9600 baud
After the 361T receives the first delimiter (#4 in the command
string), it automatically performs a "g LF" (#5 and #6 in command
string). The unit should not receive an additional "g<LF>".
A command string with a "T" as the action code with no address
preceding it, sent to the 361T at the correct baud rate will force the
361T to transmit the proper hexadecimal code for its baud rate.
U [UPDATE] command must be received by the 361T after any write
commands to internal or external RAM Memory and after any "Y"
action codes. This command forces the unit to update the data
written to RAM or process a "Y" action code and forces a position
within a function as determined by an address preceding the "U"
command. The 361T must receive the following as the address and
action code portion of the command string.
1
U———— Action Code
————— 1 = forces display to data level
0 = forces display to beginning of
function
V [VERSION] transmits the software date code (that is displayed on
the character displayed during power-up), and a <CR> or
<CR><LF> to the serial port.
X-RITE 361 VER. XXXX (XXXX = Date Code)