7. Maintenance
General
The X-Rite 361T is covered by a one-year limited warranty (excluding
lamps) and should be referred to the factory or authorized service center
for repair within the warranty period. Attempts to make repairs within
this time frame may void the warranty.
Always verify instrument calibration to assure proper instrument
operation. Make sure all connections are properly made.
X-Rite provides a factory repair service to their customers. Because of
the complexity of the circuitry all circuitry repairs should be referred to
the factory or an authorized service center.
X-Rite will repair any 361T past warranty. Shipping costs to the factory
or authorized service shall be paid by the customer and the instrument
shall be submitted in its original carton, as a complete unaltered unit.
AS A ROUTINE MAINTENANCE PROCEDURE, replace the read
lamp after every 1200 hours of operation. This allows approximately a
half year of operation (based on an 8 hour operating shift per five days a
week). Failure to do so may cause erroneous density readings. X-Rite
has also provided you with an extra Transmission read lamp. This lamp
should only be replaced by a qualified service repair person.
Exterior Cleaning
Never use any liquids (water or solvents) to clean the exterior of the
instrument. Whenever required, the exterior of the instrument can be
wiped clean with a dry, lint-free cloth.