ALIGNMENT CHECK (continued)
12. Unplug line cord then loosen two Phillips-head screws 6 in bottom
of lamp assembly.
13. Set front half of unit over edge of table, then plug line cord into wall
receptacle.
14. Carefully move lamp assembly until lamp filament is centered with
aperture (see Steps 9 and 10).
15. After alignment, unplug line cord and tighten two Phillips-head
screws [6] in bottom of lamp assembly WITHOUT moving lamp
assembly out of alignment.
16. Fasten bottom plate in position with three screws removed in
Step 2.