Operator`s manual

SETTING INSTRUMENT CONFIGURATION
4-11
2. Press the Enter # key open the
Calibration
Options
menu.
Enter Reflectances
The Enter Reflectances function is used to manually edit
the white calibration reflectance values.
To manually edit the white reflectance values:
1. Use the Tab keys $@ to highlight Enter
Reflectances.
2. Press the Enter key # to open the
Enter
Reflectances
screen.
3. Use the Tab keys $@ to scroll through reflectance values.
After desired value is highlighted, press the Enter key #
to access the
Reference menu.
4. Use the Tab keys $@ to choose the desired digit
(arrows above and below designated selection). Press
the Enter key # to access the editor.
NOTE: Highlighting Clear and pressing the Enter key # is a quick
method to zero the value.
5. Use the Tab keys $@ to highlight the desired number
and press the Enter key # to exit the editor.
6. When editing is completed, use the Tab keys $@ to
highlight Save & Exit and press Enter key #.
Enter Reflectances
White
390nm:70.02
400nm:77.52
410nm:83.06
420nm:86.43
^
<Select Values #>
Calibration Options
Full Calibration
Enter Reflectances
Cal Alert: 24 hrs
<Select Options>