CHAPTER FIVE
5-26
3. Use the Tab keys $@ to highlight the desired number
and press the Enter key # to exit the editor.
4. To select On/Off status, press the Enter key # with
Off (Murray-Davies) or On (Yule-Nielson)
highlighted.
5. When editing is completed, use the Tab keys $@ to
highlight Save & Exit and press Enter key #.
50% Dot Calibration
The 50% Dot Calibration function allows you to establish a
new "n" factor for the material by measuring a known 50%
dot patch. Black (visual density) patches should be used for
the solid and tint measurements.
To calibrate dot:
1. With the Dot Options menu displayed, press the Tab Up
key $ or Tab Down key @ to highlight "
50% Cal..."
option.
2. Press the Enter key # to access the Dot Cal menu.
3. Make sure Paper is highlighted on the display and
center the target window over paper to be measured.
4. Lower unit to target window and hold closed until
new data is displayed.
5. Make sure Solid is highlighted on the display and
center the target window over solid patch to measure.
6. Lower unit to target window and hold closed until
new data is displayed.
7. Make sure Tint is highlighted on the display and
center the target window over 50% patch to measure.
8. Lower unit to target window and hold closed until
new data is displayed.
50% DOT CALIBRATION
Paper
Solid
Tint
Save n
V 0.00
C 0.00
M 0.00
Y 0.00
<Measure Paper>