Operator`s manual

CHAPTER FIVE
5-44
Measuring/Editing a Hue/Gray Reference
The reference function is used to enter hue error and grayness
reference data into the instrument. The reference data is stored
in the instrument until changed. Hue/Gray reference values
are then compared to Hue/Gray measurements and the
difference displayed. The instrument maintains separate
reference data for each of the filter values.
To measure a reference:
1. If not selected, press the Tab Up key $ or Tab Down
key @ to highlight Reference. The reference values
displayed are the current values set in the instrument.
NOTE: Reference will not appear in the measurement list unless
HUE/GRAY–REF is selected as the active Hue/Gray mode. Refer to
Hue/Gray Measurement Mode earlier in this section.
2. Center the instrument target window over the
reference, and lower the instrument to take a reading.
The reference value(s) is updated.
To manually edit reference values:
1. If not selected, press the Tab Up key $ or Tab Down
key @ to highlight Reference. The reference values
displayed are the current values set in the instrument.
NOTE: Reference will not appear in the measurement list unless
HUE/GRAY–REF is selected as the active Hue/Gray mode. Refer to
Hue/Gray Measurement Mode earlier in this section.
2. Press the Enter key # to move the highlight to the data side
of the screen. <
Enter Ref> appears in the user dialog.
3. Press the Tab Up key $ or Tab Down key @ to highlight
desired attribute.
4. Press the Enter key # to access
Reference menu.
5. Use the Tab keys $@ to move the cursor to desired
digit. Press Enter key # to access editor.
HUE/GRAY—REF Options
Paper
Hue/Gray
Reference
h 0%
g 0%
Y —> C
<Measure Ref>
T