User's Manual
Table Of Contents
- 1 Introduction
- 2 Safety and warranty
- 3 Product description
- 4 Commissioning
- 5 XorayaSuite
- 6 Maintenance
- 7 Storage, transport and disposal
- 8 Appendix
USER MANUAL
COMMISSIONING
33
4.7 Configuring terminators
Configure the built-in terminators via the rotary switches (M) on the
back of the ML-N4000. ( Connections and controls)
For the LIN interface, select between internal or external power supply
via the toggle switch (P)
4.7.1 4x CAN
Position
Resistance
0
Open (∞)
1
2.6 kΩ
2
120 Ω
The channels 3 and 4 have the value Open and, therefore, are not
terminated.
4.7.2 7x LIN
Position
Power supply
Vi
Internal
Ve
External
Use the internal power supply of the LIN measurement card only if the
connected LIN device uses the Minilogger Z7 voltage as a reference voltage.
Bear in mind that this is not possible when using the Minilogger Z7 in a 24-V
electrical system, for example. In this case, connect the external power
supply port to the corresponding reference voltage.