- Xantrex Internal GPIB-M Interface: Multichannel Functionality for Programmable DC Power Supplies Operating Manua

Calibration
Exit calibration mode
Release 2.1 129
Exit calibration mode
When you have completed calibration, exit calibration mode by sending the
command:
CAL:STAT OFF, "0000"
where “0000” is replaced with your security code.
Restore Factory Calibration
Restore the unit to the calibration constants set at the factory by sending the
command:
CALibration:RESTore
Note This procedure should not be used in place of regular calibration, but may be useful to
restore the unit to an operational state in case of failure.