User's Manual

19
Fig. 3.2. Cross section of the dPCR cartridge. Notice the exact location of the shelf in well #D, where 40 μ
l
of 1x dPCR buffer are added.
When loading the dPCR cartridge, avoid introducing air bubbles by pipetting on the side wall of the wells.
6. Load 300 μl 1x dPCR buffer in the first well (#A). It is important to load the cartridge in the order
described here and avoid air bubbles by pipetting carefully on the side of the well.
7. Load 40 μl 1x dPCR buffer to the shelf of the collection well #D (Fig. 3.2).
8. Load 40 μl dPCR reaction mix including primers and sample DNA into well #C.
9. Load 100 μl dPCR oil into well #B.
Notice that the order of loading is: 300 μl 1x dPCR buffer in the well #A
40 μl 1x dPCR buffer to the shelf #D
40 μl 1x dPCR reaction mix in well #C
100 μl dPCR oil in well #B
Well #D shelf