Guidance and manufacturer's declaration electromagnetic immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacturers declaration – electromagnetic immunity
The Fingertip Pulse Oximeter is intended for use in the electromagnetic environment specified
below. The customer or the user of the Fingertip Pulse Oximeter should assure that it is used in
such an environment.
Immunity test
EN 60601
test level
Compliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15kV air
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lnes
N/A
Surge
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
N/A
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
< 5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(>95 % dip in U
T
)
for 5 sec
N/A