Manual

Sondes Section 2
YSI Incorporated Environmental Monitoring Systems Operations Manual 2-90
CALIBRATION RECORD THE GLP FILE
When any sensor is calibrated, 6-series sondes will automatically create a file in sonde memory that
provides details of the calibration coefficients before and after the calibration. The file will have a .glp
extension and will have the Circuit Board Serial # as the default filename. The file can be viewed by
following the path File|Directory|View from the Main sonde menu.
-------------File details--------------
1-View file
2-File:00003001.glp
3-Samples: 39
4-Bytes: 790
Filename Samples
1-BRIDGE1.dat 19
2-BRIDGE2.dat 27
3-UPLAKE.dat 33
4-CLRLAKE2.dat 167
5-DWNLAKE2.dat 31
6 - 00003001.glp 39
------------------File-----------------
1-Directory 4-View file
2-Upload 5-Quick view file
3-Quick Upload 6-Delete all files
Select option (0 for previous menu): 1