Instruction Manual

What to do if...
80
Buffer solutions:
Incorrect buffer solutions Change calibration
procedure
Buffer solutions too old Only use once.
Note the shelf life
Buffer solutions depleted Change solutions
Cause Remedy
pH electrode:
Diaphragm contaminated Clean diaphragm
Membrane contaminated Clean membrane
Sample:
pH value not stable Measure with air
excluded if necessary
Temperature not stable Adjust temperature if
necessary
Electrode + sample:
Conductivity too low Use suitable electrode
Temperature too high Use suitable electrode
Organic liquids Use suitable electrode
Cause Remedy
Batteries almost depleted Replace batteries (see
section 5.1 M
AINTENANCE)
No stable
measured value
LoBat