What to do if...
137
Cause Remedy
Batteries almost
depleted
Replace batteries
(see section 5.1.1
CHANGING THE BATTERIES)
Cause Remedy
Timeout of the interface Checkout connected
instrument
Cause Remedy
pH electrode unsuitable Use suitable electrode
Temperature difference
between buffer and
sample too large
Adjust temperature of
buffers or samples
Measuring procedure not
suitable
Follow special procedure
Cause Remedy
Operating state
undefined or EMC
electric stress unallowed
Processor reset:
Press the <auto read>
key and switch on
instrument
Cause Remedy
Printer switched off Switch on printer
No mains power supply
connected
Connect mains power
supply
Interface cable
connected
Unplug cable
No paper available Replace roll of paper
LoBat display
*( display
Obviously incorrect
measured values
Instrument does not
react to keystroke
Integrated printer
does not print