Cond 1970i Operation
43
ba75692e01 01/2007
4.6.2 Recorder (analog output)
You can transmit data to a recorder via the analog output. Connect the
analog output to the recorder via the AK323 interface cable.
The data output automatically switches to Recorder output.
Socket assignment
Note
The analog output is activated automatically in the cable by connecting
2 and 3.
The output on the analog output corresponds to the value shown on the
display.
Signal range The signal range of the analog output depends on the measured vari-
able and the measuring range:
Conductivity
Salinity
1 free
2 Plug coding
3 Ground
4 Analog output
1
2
3
4
RS 232
REC
(internal resistance < 5 Ohm)
Measuring range Voltage Resolution
0.000 ... 1.999 µS/cm 0 ... 1999 mV 1 mV
0.00 ... 19.99 µS/cm 0 ... 1999 mV 1 mV
0.0 ... 199.9 µS/cm 0 ... 1999 mV 1 mV
0 ... 1999 µS/cm 0 ... 1999 mV 1 mV
0.00 ... 19.99 mS/cm 0 ... 1999 mV 1 mV
0.0 ... 199.9 mS/cm 0 ... 1999 mV 1 mV
0 ... 500 mS/cm 0 ... 500 mV 1 mV
Measuring range Voltage Resolution
0 ... 70.0 0 ... 700 mV 1 mV